20
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
Getting started
WATER CONNECTIONS
The sterilizer clean water tank can be filled manually by the user or
automatically with a water supply system. The water supply
system must deliver demineralized or distilled water meeting the
specifications listed in these instructions. Do not add any
chemical/additive to the water.
The manufacturer’s warranty is void if the sterilizer was used with
water containing either chemical additives, or contaminant levels
exceeding those listed in these instructions. See "Feed water
specifications (EN 13060)" on page 96 and "Feed water
specifications (ANSI/AAMI and AAMI TIR34)" on page 97.
Notice: The maintenance of the external water filling system must
be done in exact accordance with the Instructions for use given with
the relevant system.
LAN CONNECTION (OPTIONAL)
1
Insert a standard Ethernet
cable to in the LAN port of
the sterilizer.
2
Insert the other end of the
cable in the LAN port of your
computer or computer
network: when the sterilizer
will be switched on it will
connect automatically to the
LAN.
WI-FI CONNECTION (OPTIONAL)
For the Wi-Fi connection proceed as follows:
1
Insert the Wi-Fi key in the USB port.
2
Read the Instructions for use provided with the Wi-Fi key.