8
MN-111 | Service book | Errors and Troubleshooting MN-111 ENG Rev08 | 2019/06/21 | © 2019 W&H Sterilization Srl
Safety warnings
GENERAL WARNINGS
n
Before performing any repairing or service interventions on the sterilizer, ALWAYS DISCONNECT the electric power
supply cable to avoid injury.
n
If for any reason you are unable to remove the power supply or if you have to walk away from the unit, always
place a “WORK IN PROGRESS” sign, to warn other people that might walk up to the sterilizer.
n
Service should be performed by qualified technicians only.
n
Observe all local and national directives related to service and safety standards.
n
Always use personal safety devices (e.g. glasses, gloves, protection mask, etc.).
n
Before making any repairs or transport, ward off unauthorized personnel from the working area.
n
Refer also to all the safety indications given in the Instructions for use of the product.
TRANSPORT
Spare parts shall be handled with care.
In case of transport of the sterilizers:
n
Completely drain both water tanks;
n
Drain the condenser bottle;
n
Allow the sterilization chamber to cool down;
n
Use original or appropriate packaging;
n
Put the box of the sterilizer on a desk
n
Remove the upper part of the box;
n
Position the sterilizer in the dedicated area of the box;
n
Carefully seal the box;
n
The transport of the sterilizer shall be done by two authorized technicians using PPE devices (personal protective
equipment) according to applicable standards;
n
The sterilizer shall be handled with care.