LISA | Instructions for Use | VA131 2019 ENG Rev06 | 16/04/2020 | © 2020 W&H Sterilization Srl
21
Getting started
INSTALLING THE STERILIZER
WARNING! In case of sterilizer malfunctions immediately
unplug the sterilizer and call for service. Do not attempt to
repair the sterilizer by yourself.
Notice:
Please ensure that all installation requirements are met before
plugging the sterilizer. See "Connection diagrams" on page 125.
No other devices should be connected to the sterilizer power panel
circuit.
1
Place the sterilizer on a sturdy, flat and level surface.
2
Remove all items from the sterilizer chamber except the
chamber rack. Remove all plastic covers from trays.
3
Connect the auto-fill and auto-drain tubes in the rear of the
sterilizer.
4
Connect the Ethernet cable or the Wi-Fi key in the rear of the
sterilizer.
5
Attach the power cord to the socket in the rear of the sterilizer
and route the cord through the cable guides.
6
Connect the power cord to a wall outlet. For power supply
requirements, see "Technical data" on page 123.
Operating the sterilizer
POWER THE STERILIZER ON/OFF
1
Press the power switch behind the
service door: once switched ON, the
visual indicator on the power
switch will turn green.