Technical data
86
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
Unwrapped
Wrapped Medium
Load
Wrapped Large
Load
Wrapped XL Load
Low Temperature
Textiles
No
No
No
No
Yes
Maximum load weight (excluding trays)
instruments:
18 lb (8.1 kg)
instruments:
3 lb (1.4 kg)
instruments:
6 lb (2.7 kg)
instruments:
14 lb (6.3 kg)
included cassettes
instruments:
5 lb (2.2 kg)
included trays
or textile:
4.4 lb (2 kg)
1
: the total cycle time may vary depending on the type of load (solid or porous), the load weight, and other factors.
Sterilization program phases
LEGEND OF THE STERILIZATION PROGRAM PHASES
Following is the description of the sterilization phases.
Code
Phase
Description
PHE
Pre-heating
phase
Pre-heating of the sterilizer steam generator and chamber. This phase is not considered a part of the cycle.
HEA1
HEA2
Heating
phase (steam
pressure
pulses)
Initial heating phase to remove the air from the sterilization chamber. Air is removed using pressure pulses followed by discharge phases. The
iteration of these phases reduce the percentage of air inside the chamber and determinates the correct sterilization conditions.
HEA3
Pressure rise
Heating phase to achieve the sterilization conditions (pressure and temperature) for the selected cycle.