Maintenance
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
63
1200-cycle or yearly maintenance
REPLACING THE DOOR GASKET
Notice
: the door gasket needs to be replaced every 1200 sterilization cycles or once a year, whichever come first. A replacement message alerts
when replacement is due. If the consumables is replaced prior to the message, you have to reset the consumable cycle counter.
Proceed as follows:
1
Open the chamber door.
2
Remove the used door gasket by hand.
3
Carefully clean the seal seat and the inside face of the chamber door.
Note
: tap
to see the animated replacement procedure.