52
Lexa | Instructions for Use | MN-111 Med ENG Rev14 | 03/04/2023 | © W&H Sterilization Srl
Sterilization cycles
END OF A STERILIZATION CYCLE
When a cycle is successfully finished, the "Cycle completed"
message appears on the screen. To end the cycle:
1
Tap
to view the cycle summary or tap
to view the cycle
parameters. See "View the cycle parameters" on the previous
page.
2
Tap
OPEN
to open the door: the door unlocks and the home
page appears.
Note
: if an error message appears see "Troubleshooting" on page 73
CAUTION!
Hot surfaces.
Burnings. Do not touch the
chamber, the internal side
of the door and the internal
fittings. Use the tray holder
or cassette holder or
gloves for high
temperatures or adequate
protection to remove the
load!
3
Open the chamber door.
4
Remove the load and stock it.
STOP A STERILIZATION CYCLE
WARNING! You can stop the cycle at any time. Instruments
must not be considered sterile if this occurs before the
DRY phase.
A cycle can be manually aborted at any time. To stop a cycle:
1
Tap
STOP
: a confirmation request appears.
2
Tap
to abort the stop command. The cycle continues
programmed.
3
Tap
to abort the cycle: the sterilizer starts a reset phase.
Notice
: do not switch off the sterilizer during the reset phase: it
takes some time to reset the system and reach safe conditions in
the sterilizer chamber.
4
Check the message. See "Messages of a stopped sterilization
cycle" on the next page.
5
Tap
to view the cycle parameters. See "View the cycle
parameters" on the previous page.
6
Open the chamber door.
7
Reprocess the load if necessary.
CAUTION!
Hot steam. Wait the steam to dissipate before
opening the door.
Summary of Contents for MN-111
Page 1: ...MN 111 Instructions for Use MN 111 Med ENG Rev14...
Page 2: ......