Maintenance
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
65
CLEANING THE CHAMBER FILTER
Proceed as follows:
1
Allow the sterilization chamber to cool down.
2
Empty the sterilizer chamber by removing the trays and the rack.
3
Turn the filter cap at the back of the chamber (bottom/center) counter-clockwise.