User Interface
31
Waltron User Manual
3051 Hardness Analyzer
5.3
MAIN SCREEN
The main screen displays:
Name of the analysis program (Hardness Analyzer)
Login **** window (two levels of password); pressing on **** the user can input the 4
number password
Buttons that allow access the sub-menus (Run, Display, Program, Service and ?)
Value (ppm or ppb) of the last analysis
Alarms / relays
Active step of the analysis in progress
Reagent(s) consumption (%)
Table 5.3: Analyzer operating statuses.
ANALYZER OPERATING STATUS
STAND-BY
The analyzer is waiting for a user’s command (for example when the analyzer
is switched on after a shutdown), after single cycle analysis, or when the
operator has forced a stop in stand-by at the end of the current online / single
cycle by pressing the button previously activated a second time.
SINGLE
CYCLE
The analyzer is performing a single analysis cycle. As soon as the cycle is
completed the analyzer will go into the stand-by condition.
ON-LINE
The analyzer is performing a continuous cyclic analysis based on the steps
set in the PROGRAM menu (Analysis Cycle) and started when the START
ON LINE button has been pressed. As soon as the analyzer completes an
analysis cycle, it will restart with a new analysis. This condition is identified as
the online condition.
MANUAL
STOP
The analyzer has been forced to stop. A message in the main menu shows
the “manual stop” and the fault alarm is initialized.
Table 5.4: Main menu button descriptions.
MAIN MENU BUTTONS
RUN
This menu allows the user to put the analyzer into online measurement mode
or run a single analysis cycle on stream 1 or stream 2. The user can also
stop the analyzer with the emergency stop button.
DISPLAY
This menu displays all analysis data, the active step of the current analysis
cycle with the time counter, the sensor's signal trend graph, and allows
access to the manual step functions.
PROGRAM
This menu allows access to the calibration menu, enabling and disabling the
loss of sample alarm, the 4-20 mA output, and settings of the analyzer.
SERVICE
Second level admin password required. This menu allows access to the
analysis parameter, time, blank value, and regulation of the 4-20 mA output,
led, and reference.
?
This menu shows the help user guide.
Summary of Contents for AQUALERT Aqualyzer 3051
Page 52: ...Appendix A Panel Mount Schematics 51 Waltron User Manual 3051 Hardness Analyzer ...
Page 53: ...Appendix B Help Menu 52 Waltron User Manual 3051 Hardness Analyzer APPENDIX B HELP MENU ...
Page 54: ...Appendix B Help Menu 53 Waltron User Manual 3051 Hardness Analyzer ...
Page 55: ...Appendix B Help Menu 54 Waltron User Manual 3051 Hardness Analyzer ...
Page 56: ...Appendix B Help Menu 55 Waltron User Manual 3051 Hardness Analyzer ...
Page 57: ...Appendix B Help Menu 56 Waltron User Manual 3051 Hardness Analyzer ...
Page 58: ...Appendix B Help Menu 57 Waltron User Manual 3051 Hardness Analyzer ...
Page 59: ...Appendix B Help Menu 58 Waltron User Manual 3051 Hardness Analyzer ...
Page 60: ...Appendix B Help Menu 59 Waltron User Manual 3051 Hardness Analyzer ...
Page 61: ...Appendix B Help Menu 60 Waltron User Manual 3051 Hardness Analyzer ...
Page 62: ...Appendix B Help Menu 61 Waltron User Manual 3051 Hardness Analyzer ...