SPECIFICATIONS
1.1.1
PERFORMANCE
Sensor Classification:
Cationic Exchange
Application:
Measurement of specific, cation, and degassed cation
conductivities in high purity process water
Measuring Range:
0 to 9.999 µS/cm, 0 to 99.99 µS/cm
Accuracy:
< 0.6% of the measuring range
Degassing Efficiency:
93.5%
User Interface:
5.5” color touch screen
Data Storage:
Extract via USB 2.0
Degree of Protection:
Transmitter IP67, Corrosion Resistant Finishes
Materials in Contact with
Sample:
Stainless steel, PVC
Power Supply:
AC 110 to 240 V +10 / -15%; 48 to 63 Hz or AC/DC 20
to 30 V; 48 to 63 Hz
Required Maintenance:
Replace Cation Resin Bottle (monthly)
Replace Decarbonization Column (monthly)
Mounting:
Wall or Panel Mount
Dimensions:
29 x 37 x 10” (613 x 939 x 254 mm)
1.1.2
OPERATING CONDITIONS
Ambient Temperature:
32
– 140° F (0 – 60 °C)
Sample Medium:
Liquid
Sample Temperature:
32
– 122° F (0 – 50 °C)
Sample Flow Rate:
.031 - .044 GPM (7-10L/h)
Sample Pressure:
Maximum of 30 psig (2 bar)
Outlet Sample Pressure:
Atmosphere
1.1.3
COMMUNICATIONS & SIGNAL OUTPUTS
Analog Outputs:
Four 0/4
– 20 mA for measured signals
Serial I/O for Signals:
RS422/RS485 (optional)
Alarms:
General Alarm, Low Degassing Efficiency Alarm, Resin
Exhausted Alarm
Optional: