Analyzer Initial Start-Up
25
Waltron User Manual 101-046-B.3
9096 Degassed Cation Conductivity Analyzer
4.1.2
INSTALLING DECARBONIZATION COLUMN
Step
Illustration
Description
1
Note the date and time that you
are installing the cartridge.
2
Remove plugs covering the
inlet (bottom or cartrdige) and
outlet (top of cartridge) holes of
the new cartridge.
3
Screw the new cartridge into
place.
4.2
PUTTING THE ANALYZER ONLINE
Once the analyzer has been installed, electrical connections, and sample connections have been
made, the analyzer is ready for online measurement.
Turn on sample flow to the analyzer and adjust to 8-10L/hr (130-170ml/min). The analyzer will
begin to display measurement readings for the specific, cation, and degassed cation
conductivities, as well as the calculated pH. It will take about 1 hour of online sample flow until
the measurements are accurate. In order for the degassing column to operate correctly, a small
amount of sample flow must discharge contantly from the overflow chamber into the drain cup.
It is important to note that if Waltron’s pre-rinsed resin is not used, it may take anywhere from
several hours to several days for the cation and degassed cation conductivities to read correctly.