![Waltron 9031CX User Manual Download Page 51](http://html1.mh-extra.com/html/waltron/9031cx/9031cx_user-manual_3304562051.webp)
Specifications
50
Waltron User Manual 101-045-E.1
9031CX Sodium Analyzer
7
SPECIFICATIONS
Range
0.1ppb
– 10ppm
Accuracy
+/-
5% of reading or +/- 0.1ppb (whichever is greater) within
+/-5C of calibration temperature
Reproducibility
+/-
5% of reading or +/- 0.1ppb (whichever is greater) at
constant temperature
Response Time
90% of 1-10ppb step: 4 minutes; 90% of 100-1ppm step: 5
minutes
Current Outputs
Two isolated 4-20mA current (analog) outputs
Alarms
Three voltage-free contacts, alarm points set from
transmitter rated at 2A, 250VAC (non-inductive)
Power
Wide range of input power supply 90VAC
– 250VAC
Sample
Temperature: 41-131F (5-55C); Flow 150-400ml/min
Ambient Temp
32-131F (0-55C)
Pressure
5-30psig
Composition
Sample should be filtered to at least 60 microns, free of
film forming compounds
Power Consumption
Less than 20VA
Sample Inlet Fitting
1/4” Swagelok
Sample Outlet Fitting
Nozzle that fits 3/8” hose connection