EC     REP

6

7

Customization 

(

Note

: Customization should only be performed by your healthcare provider. If using the HEELIFT® Ultra 

boots pull back the removable elevation cover first )

The HEELIFT® offloading boot can be customized to offload other at-risk areas in addition to 
the back of the heel including the Achilles Tendon, Malleolar Decubitus, Foot Drop, hip and/or 
foot rotation, and SCD Tubing.  

1. Achilles Tendon. 

(see exhibit K.)

• Make an upside-down V cut or an upside-down U cut as determined 

appropriate by your healthcare provider in the bottom of the fixed 
elevation pad to appropriately offload the Achilles tendon.

2. Malleoli (ankle bones) 

(see exhibit L.)

• Make an oblique cut or an upside-down hockey stick cut as 

determined appropriate by your healthcare provider on the side of 
the elevation pad starting from a point just above the ankle bone 
and cutting downward through the bottom of the fixed elevation 
pad so that the ankle bone is completely offloaded. 

• If further offloading is required as determined by your healthcare 

provider place the extra foam pad in the groove on the side of the 
ankle bone needing to be offloaded with the bottom of the pad just 
above the ankle bone and cut the top of the extra foam pad flush with the 
top of the boot.  Attach the extra foam pad to the side of the boot once appropriately 
positioned in the groove by peeling the adhesive backing off the extra foam pad. Make sure the 
adhesive side never touches the skin. 

3. Foot Drop 

(see exhibit M.)

• Measure the extra foam pad from the tip of the toes to the area 

where the arch meets the heel.

• Cut the extra foam pad at the area where the arch meets the heel 

and bevel the edge.

• Peel the adhesive backing off the extra foam pad.
• Position the extra foam pad vertically under the patient’s foot so that 

the beveled edge is under the area where the arch meets the heel, and 
the adhesive side of the pad, is against the bottom of the boot to not touch the skin (make sure 
adhesive never touches the skin). 

4.Hip and/or Foot Rotation 

(see exhibit N.)

• Use the optional Anti-Rotation Wedge to prevent hip and/or foot 

rotation. See instructions included with the wedge.

• The Anti-Rotation Wedge gives you the flexibility needed to stop 

rotation in immobile patients. The easy to apply, reusable wedge 
attaches to either side of the boot, and prevents rotation.

• Works for both external and internal rotation.

5.SCD Tubing Placement 

(see exhibit O.)

• Place SCD tubing in the groove built into the boot along the side of 

the boot, between the elevation pad and the side of the boot 
opposite the side with the attached D-rings. The allows the tubing to 
lay flat and to exit the boot without touching the skin.

• SCD tubing should always carefully be placed into the groove, built 

into the boot, so that it never touches the skin to avoid compromise 
of the skin from the tubing.

• Ventilation holes are meant for air circulation within the boot.  Do not 

place any tubing through the ventilation holes in the boot.  

Cleaning 

(

Caution

: Clean or disinfect using a nonbleach gentle detergent or a nonbleach disinfectant.  As with any medical 

device, bleach can breakdown the integrity of the material of the HEELIFT® offloading boot.  To maintain the integrity of the HEELIFT® we 
recommend sanitizing or hand washing.)

1. To sanitize

• Use a nonbleach disinfectant wipe or spray.

2. To hand wash

• Wash in warm water with a mild nonbleach detergent.

3. To machine wash and dry

• Secure all straps.
• Remove traction device is using HEELIFT® traction boot.
• Place in mesh HEELIFT® brand, regular mesh laundry bag, or pillowcase.
• Machine wash in warm water (60˚C/140˚F) gentle cycle with a mild nonbleach detergent.
• Machine dry on low heat.    

MADE IN CHINA

K.

L.

M.

N.

O.

4 To sterilize

• Keep the straps loosely affixed and 

sterilize in steam    autoclave using 
standard protocol.

• Allow the HEELIFT® boot to completely 

cool and dry before reapplying device.

Walgreen Health Solutions, LLC
1316 Sherman Avenue
Evanston, IL 60201
Phone: (847) 328-9540
Email: [email protected]
www.walgreenhealthsolutions.com

Distributor
Moorings Medical 
Equipment UK Ltd.
51 Slaght Road
BT42 2JH N. Ireland
Phone: 0044(0)28 2563 7777

Emergo Europe B.V.
Prinsessegracht 202514AP
The Hague, Netherlands

Dutch

French

German

Italian

Lithuanian

Norwegian

Polish

Portuguese

Romanian

Slovak

Slovenian

Spanish

Turkish

Reviews: