RELATED FUNCTIONS
22
TC6XX-TM6XX Series
D.3
Calibrated sensors
The calibrated sensors function of the device makes it possible to use sensors, the calibration (correction) coefficients of
which are taken into consideration by the device during measurement.
Using the F1 key, enter the Configuration menu.
Select the
Calibrated Sensors
function, followed by one of the 5 available sensors.
Confirm (
VAL
key).
Enter the sensor information fields. Use the F1 function key (
) to move from one field to the next.
Confirm your choice using the
VAL
key.
To add a value in the table of calibration points, use the
keys, enter the calibration points (real value and
value read) then confirm using the VAL key.
Repeat this operation for all the calibration points (maximum of 4).