7-2
Chapter 7: Operational Verification Procedure
33373-001 Version A
7.1.4 Select LCD Test
1. Press the Extended Functions Icon. Press Vent Set Up Icon.
2.
Select “Dim Screen” and verify screen intensity level dims
3.
Select “Bright Screen” and verify the screen returns to normal intensity
4.
Select “Video Normal” and verify Screen has turned white
5.
Select “Video Inverse” and verify the screen returns to normal operation
6. Press the Exit icon on the touch screen.
7.1.5 Leak Test
1. Connect a non-disposable test circuit and test lung to the unit. Completion of the test takes
about 15 seconds
2. Press the Leak Test Icon. The test lung should inflate and hold the pressure for approximately
10 seconds.
3. Verify test passes.
NOTE
If any test does not pass, do not continue with performance test. Turn to Chapter 9
“Troubleshooting” section under
“UVT Test”
Tests and follow the troubleshooting guide for the
failed test.
7.1.6 FiO2 Monitor Calibration
1. Press the Extended Functions Icon.
2. Press the Vent Set Up icon and verify the Fio2 Monitor is Enabled. Press the Exit icon.
3. Press the FiO2 Mon Calibration Icon.
4. Select the Start Ambient Calibration icon and verify the calibration has completed and is valid.
This calibration can take up to 3
–5 minutes.
5. Select the Start 100% Calibration icon and verify the calibration has completed and is valid.
6. Press the Exit icon.
7. On the UVT screen press the Exit icon and verify the unit restarts in regular operating mode.
8. Disconnect the Leak Test circuit and connect a suitable permanent Patient Test Circuit
(Do not
use disposable circuits for testing purposes)
using the Siemens test lung P/N 33754 to continue
the “Performance Tests”
7.2 Performance Test
1. Ensure that a 50 PSIG Oxygen gas source is connected to the High Pressure inlet assembly on
the rear of the VELA. Connect the VELA power cord to a properly grounded AC outlet.
2. Set Ventilator parameters as follows:
Mode:
Volume Assist Control
Volume:
500 ml
Peak Flow
20 lpm
Breath Rate
12 bpm
PEEP
0 cmH
2
O
3. Adjust all other controls and alarm limits out of range of operation to
either “OFF” or min/max
settings for alarms. .
Summary of Contents for VELA
Page 2: ...33373 001 Version A VELA Ventilator Service Manual ...
Page 19: ...2 6 Chapter 2 Theory of Operation 33373 001 Version A This page intentionally left blank ...
Page 87: ...B 2 Appendix B Reordering Instructions 33373 001 Version A This page intentionally left blank ...
Page 88: ...Appendix C Schematics Diagrams C 1 33373 001 Version A Appendix C Schematics Diagrams ...
Page 89: ...Appendix C Schematics Diagrams C 2 33373 001 Version A ...
Page 90: ...Appendix C Schematics Diagrams C 3 33373 001 Version A ...
Page 91: ...Appendix C Schematics Diagrams C 4 33373 001 Version A ...
Page 92: ...Appendix C Schematics Diagrams C 5 33373 001 Version A ...
Page 93: ...Appendix C Schematics Diagrams C 6 33373 001 Version A ...
Page 94: ...Appendix C Schematics Diagrams C 7 33373 001 Version A ...
Page 95: ...Appendix C Schematics Diagrams C 8 33373 001 Version A ...
Page 96: ...Appendix C Schematics Diagrams C 9 33373 001 Version A ...
Page 97: ...Appendix C Schematics Diagrams C 10 33373 001 Version A ...
Page 98: ...Appendix C Schematics Diagrams C 11 33373 001 Version A ...
Page 99: ...Appendix C Schematics Diagrams C 12 33373 001 Version A ...
Page 100: ...Appendix C Schematics Diagrams C 13 33373 001 Version A ...
Page 101: ...Appendix C Schematics Diagrams C 14 33373 001 Version A ...
Page 102: ...Appendix C Schematics Diagrams C 15 33373 001 Version A ...
Page 103: ...Appendix C Schematics Diagrams C 16 33373 001 Version A ...
Page 104: ...Appendix C Schematics Diagrams C 17 33373 001 Version A ...
Page 105: ...Appendix C Schematics Diagrams C 18 33373 001 Version A ...
Page 106: ...Appendix C Schematics Diagrams C 19 33373 001 Version A ...
Page 107: ...Appendix C Schematics Diagrams C 20 33373 001 Version A ...
Page 108: ...Appendix C Schematics Diagrams C 21 33373 001 Version A ...
Page 109: ...Appendix C Schematics Diagrams C 22 33373 001 Version A ...