1.3 Recommended Maintenance Schedules
1.3.1 Schedules
Every 500 hours,
the fan and ambient air filters should be cleaned and replaced if necessary.
Every year,
the following preventive maintenance procedure should be performed (see chapter 4
for instructions). This procedure includes:
Install PM kit P/N 11416
Replace the 40 psi Regulator and the Blender Screws
Replace the Internal Oxygen Sensor, P/N 27750-001 (Analytical Industries) or 27751-001
(Maxtec)
Perform verification procedures described in Chapter 4
Calibration of the transducers & solenoids if necessary.
Every
two years
, internal batteries P/N 16048, includes the battery tray should be replaced along
with the annual PM parts and procedures.
Maintenance on the VELA should only be carried out by a Vyaire trained and authorized service
technician. Vyaire will make available to qualified technicians, service manuals and such items as
circuit diagrams, component parts lists, calibration instructions and other information to assist in
repair of those parts of the ventilator designated by the manufacturer as repairable items.
The drawings, diagrams and schematics included in this manual are for reference only and may be
updated separately from this manual after publication. For current revisions of all documentation,
contact Vyaire, Technical Support at the numbers provided in Appendix A.
Summary of Contents for VELA
Page 2: ...33373 001 Version A VELA Ventilator Service Manual ...
Page 19: ...2 6 Chapter 2 Theory of Operation 33373 001 Version A This page intentionally left blank ...
Page 87: ...B 2 Appendix B Reordering Instructions 33373 001 Version A This page intentionally left blank ...
Page 88: ...Appendix C Schematics Diagrams C 1 33373 001 Version A Appendix C Schematics Diagrams ...
Page 89: ...Appendix C Schematics Diagrams C 2 33373 001 Version A ...
Page 90: ...Appendix C Schematics Diagrams C 3 33373 001 Version A ...
Page 91: ...Appendix C Schematics Diagrams C 4 33373 001 Version A ...
Page 92: ...Appendix C Schematics Diagrams C 5 33373 001 Version A ...
Page 93: ...Appendix C Schematics Diagrams C 6 33373 001 Version A ...
Page 94: ...Appendix C Schematics Diagrams C 7 33373 001 Version A ...
Page 95: ...Appendix C Schematics Diagrams C 8 33373 001 Version A ...
Page 96: ...Appendix C Schematics Diagrams C 9 33373 001 Version A ...
Page 97: ...Appendix C Schematics Diagrams C 10 33373 001 Version A ...
Page 98: ...Appendix C Schematics Diagrams C 11 33373 001 Version A ...
Page 99: ...Appendix C Schematics Diagrams C 12 33373 001 Version A ...
Page 100: ...Appendix C Schematics Diagrams C 13 33373 001 Version A ...
Page 101: ...Appendix C Schematics Diagrams C 14 33373 001 Version A ...
Page 102: ...Appendix C Schematics Diagrams C 15 33373 001 Version A ...
Page 103: ...Appendix C Schematics Diagrams C 16 33373 001 Version A ...
Page 104: ...Appendix C Schematics Diagrams C 17 33373 001 Version A ...
Page 105: ...Appendix C Schematics Diagrams C 18 33373 001 Version A ...
Page 106: ...Appendix C Schematics Diagrams C 19 33373 001 Version A ...
Page 107: ...Appendix C Schematics Diagrams C 20 33373 001 Version A ...
Page 108: ...Appendix C Schematics Diagrams C 21 33373 001 Version A ...
Page 109: ...Appendix C Schematics Diagrams C 22 33373 001 Version A ...