Caution
Make sure the pressure relief valve on any LN
2
tank is adjusted to
30 PSI maximum blow-off.
s
Warning
Carbon dioxide gas suppresses oxygen levels and may cause
suffocation if area is not well ventilated. Refer to “Handling Liquid CO
2
in Appendix B of this manual.
s
Power
- indicates the unit has AC power.
1. Normal Operation
•
Low Battery: Illuminates only after a battery test which occurs every 8
hours. The battery test is a loaded test and during the test, the battery
voltage is monitored. When the battery voltage is less than a certain
threshold, this light will illuminate and stay illuminated until the next
battery test.
• Solenoid Engaged: Illuminates any time the controller is activating the
solenoid. Illumination will only occur when the controller senses the
solenoid is active and calling for injection.
2. Fault Code
This same user interface will also permit fault codes to be easily viewed
when the user presses and releases the Press to Test button in less than
2 seconds. When this occurs, both the Low Battery LED and Solenoid
Engaged LED will blink twice ON/OFF quickly indicating that fault
code display mode is active. For the next ~10 seconds, the fault codes
will be displayed as shown in the table below. The display shows only
the lower number fault code until that fault is resolved.
Chest Freezer
7-5
VWR International
Section 7
Factory Installed Options
BUS Control Panel
(cont)
Low Battery LED
Solenoid Engaged LED
Fault Code
Number
Fault Code Definition
OFF
OFF
0
Normal Operation
OFF
Blinks 1 time
1
Primary Probe Failure
OFF
Blinks 2 times
2
Secondary Probe Failure
OFF
Blinks 3 times
3
Solenoid Driver Failure
OFF
Blinks 4 times
4
Low Tank
Figure 7-6.
BUS Control Panel
Table 7-1.
BUS Fault Codes
Summary of Contents for VWRC0386A
Page 55: ...8 1 VWR Collection VWR International Section 8 Parts ...
Page 56: ...VWR Collection 8 2 VWR International Section 8 Parts ...
Page 59: ...8 5 VWR Collection VWR International Section 8 Parts ...
Page 60: ...VWR Collection 8 6 VWR International Section 7 Specifications ...
Page 61: ...VWR Collection 8 7 VWR International Section 8 Parts ...
Page 68: ...9 1 VWR Collection VWR International Section 9 Refrigeration Schematics ...
Page 69: ...VWR Collection 9 2 VWR International Section 9 Refirgeration Schematics ...
Page 70: ...VWR Collection 9 3 VWR International Section 9 Refirgeration Schematics ...
Page 71: ...9 4 VWR Collection VWR International Section 9 Refrigeration Schematics ...
Page 72: ...10 1 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 73: ...VWR Collection 10 2 VWR International Section 10 Electrical Schematics ...
Page 74: ...10 3 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 75: ...VWR Collection 10 4 VWR International Section 10 Electrical Schematics ...
Page 76: ...VWR Collection 10 5 VWR International Section 9 Section title ...
Page 77: ...10 6 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 78: ...VWR Collection 10 7 VWR International Section 10 Electrical Schematics ...
Page 79: ...10 8 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 80: ...VWR Collection 10 9 VWR International Section 10 Electrical Schematics ...
Page 81: ...10 10 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 82: ...VWR Collection 10 11 VWR International Section 10 Electrical Schematics ...
Page 83: ...10 12 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 93: ...VWR International D 2 Chest Freezer ...
Page 94: ......