6 | Operation

TD22

16 / 27

B-H86.0.0D.DB201-1.0

L,TE

OFF

Backlighting deactivated

15 30 60 120 240

Automatic shut-off of the backlighting after a selec-
ted time in seconds, during which no buttons have
been pressed

ON

No automatic shut-off of the backlighting

Temperature unit

UN,T

°

(

Temperature display in °C

°

F

Temperature display in °F

Factory settings

IN,T

NO

Use current configuration

YES

Reset product to factory settings. 

IN,T DONE

 appears

in the display

Outcome of an action

The changed value is saved and the 

Configuration

 menu is closed. 

STOR

 appears in

the display. If necessary, the product is restarted automatically in order to adopt the
changed values.

 NOTE

The configuration is closed if no button is pressed for 2 minutes. Any changes made
up to that point are not saved. 

C.END

 appears in the display.

6.2.4 Adjustment of the measuring input

Description

The temperature input can be adjusted with the zero point correction and the gradient
correction. If an adjustment is made, you change the pre-adjusted factory settings.
This is signalled with the 

T.OF

 or 

T.SL

 when the product is switched on. The standard

settings of the zero point value and the gradient value is 

0.00

. It signals that no correc-

tion is made.

In order to adjust the product, you must first open the 

Adjustment

 menu. The menu is

opened as shown in the illustration.

Prerequisites

– Sufficiently full batteries are inserted in the product.

– The product is switched off.

– Ice water, regulated precision water baths or a water bath with a reference meas-

urement are available as a reference.

Instruction

1. Press and hold the 

Down button

.

2. Press the 

On/Off button

 to switch on the product and open the 

Configuration

menu. Release the 

Down button

. The display shows the first parameter.

3. By briefly pressing the 

Function key

, you can scroll through the parameters. Select

the parameter you would like to configure.

4. When you have selected the desired parameter, change the parameter to the de-

sired value with the 

Up button

 and the 

Down button

.

5. In order to save the new parameter value, press and hold the 

Function key

 for

longer than 1 second.

Summary of Contents for TD22

Page 1: ...EN TD22 Alarm thermometer Operating manual European catalogue number EU connection 620 2643 Version 1 0 Issued 07 01 2019...

Page 2: ...product at a glance 12 5 1 The TD22 12 5 2 Display elements 12 5 3 Operating elements 12 6 Operation 14 6 1 Commissioning 14 6 1 1 Explanation 14 6 2 Configuration 14 6 2 1 Explanation 14 6 2 2 Openi...

Page 3: ...Table of contents B H86 0 0D DB201 1 0 3 27 10 Disposal 23 11 Technical data 24 12 Service 25 12 1 Online resources 25 12 2 Sales partners 26...

Page 4: ...ress of the manufacturer TD22 4 27 B H86 0 0D DB201 1 0 1 Legal address of the manufacturer VWR International bvba Researchpark Haasrode 2020 Geldenaaksebaan 464 B 3001 Leuven Phone 32 16 385011 http...

Page 5: ...m material and production de fects for a period of two 2 years from the time of delivery If there is a defect VWR shall decide on its own discretion whether to repair or replace the product free of ch...

Page 6: ...each step Instruction Tasks to be carried out by the personnel user are represented as numbered instruc tions Adhere to the sequence of the specified instructions Representation Shows an illustrative...

Page 7: ...0 7 27 NOTE For information about the software version press and hold the ON button to switch on the product for longer than 5 seconds The series is shown in the main display and the software version...

Page 8: ...s and the device specific safety instructions for this document are observed If these notices are disregarded personal injury or death as well as property damage can occur DANGER Incorrect area of app...

Page 9: ...probes with care Fit a protective cap on the measuring probe NOTE If the product is stored at a temperature above 50 C or is not used for an extended period of time the batteries must be removed Leaks...

Page 10: ...rements for which purpose this document makes a valuable contribution The instructions in this docu ment must be understood observed and followed In order to ensure that no risks arise from the interp...

Page 11: ...ude the dust proof and waterproof design in accord ance with IP 65 67 and the 3 line illuminated display which offers overhead display at the push of a button The product can be switched on switched o...

Page 12: ...hold Auxiliary display Measurement of the current temperature in min max hold mode with unit 5 3 Operating elements On Off button Press briefly Switch on the product Activate deactivate lighting Long...

Page 13: ...TD22 The product at a glance 5 B H86 0 0D DB201 1 0 13 27 Long press 2s Open menu frozen measurement is displayed Close menu changes are saved...

Page 14: ...roduct for your purposes NOTE There are various configuration parameters available depending on the product ver sion and configuration They can differ depending on the product version and configur ati...

Page 15: ...eter you would like to configure 2 Adjust the desired configuration in the selected parameter with the Up button and Down button 3 The available configuration options are listed for each parameter in...

Page 16: ...ection If an adjustment is made you change the pre adjusted factory settings This is signalled with the T OF or T SL when the product is switched on The standard settings of the zero point value and t...

Page 17: ...elected parameter with the Up button and Down button 3 The available configuration options are listed for each parameter in the following representation Representation Parameter Values Meaning Zero po...

Page 18: ...setpoint gradient correction display gradient correction T OF 1 100 T SL 37 0 C 36 6 C 0 2 1 100 0 54 Outcome of an action The changed value is saved and the Configuration menu is closed NOTE If the p...

Page 19: ...ntal temperature influ ence the measurement result NOTE Thermally conductive paste between the measuring sensor and surface can also in crease measurement accuracy in some cases 7 1 3 Cooling evaporat...

Page 20: ...Battery indicator If the empty frame in the battery display blinks the batteries are depleted and must be replaced However the device will still operate for a certain length of time If the BAT displa...

Page 21: ...requisites The product is switched off A suitable PH1 is available Instruction 1 Unscrews the Phillips screws and remove the cover 2 Carefully replace the two Mignon AA batteries Ensure that the polar...

Page 22: ...m error Product is defective Battery depleted Error in the product Product is defective Replace battery Send in for repair BAT Battery depleted Battery depleted Replace battery ERR 1 Measuring range e...

Page 23: ...or recycling It is also your re sponsibility to decontaminate the equipment in the event of biological chemical and or radioactive pollution in order to avoid putting any persons assigned with the dis...

Page 24: ...ut kink protection Weight 150 g incl battery and sensor Operating conditions 20 to 50 C 0 to 95 r h temporarily 100 r h Storage temperature 20 to 70 C Current supply 2 AA battery included in the scope...

Page 25: ...te at www vwr com All contact information of the technical customer service VWR online catalogue and information about accessories and compatible products Additional product information and special of...

Page 26: ...el 36 52 521 130 Email info hu vwr com India VWR Lab Products Private Limited No 139 BDA Industrial Suburb 6th Main Tumkur Road Peenya Post Bangalore India 560058 Tel 91 80 28078400 Email vwr_india vw...

Page 27: ...il info fr vwr com 0 18 TTC min Netherlands VWR International B V Postbus 8198 1005 AD Amsterdam Tel 31 020 4808 400 Email info nl vwr com Norway VWR International AS Haavard Martinsens vei 30 0978 Os...

Reviews: