vwr.com I Instruction manual VWR
®
CO
2
incubator
LAN port
11.5.
Each unit can be connected to the Ethernet network or directly to the computer with a LAN cable. Premium Control software
enables to create programs and send them to the unit remotely, read data (saved data and event log), create graphs and make
reports. With this feature, equipment can be controlled and monitored remotely. It is also possible to connect several units
at the same time and control them from one computer.
Temperature protection
12.
Temperature protection is included as standard. If damage occurs to the temperature controller or the user changes
the temperature settings outside the limit, temperature protection will activate.
Protection class 3.1 is a standard. The figure below shows how this works.
Protection Class 3.1 according to DIN 12880 norm is called a protecting sample function. The user can set the temperature
protection (upper limit) by himself and when the set temperature falls outside of this range, the power heater is switched off.
When the temperature returns to within normal parameters, the equipment continues to work normally.
Remote control of the unit
13.
Every Premium unit can be connected to the Ethernet network and controlled remotely using the Premium Control software.
The features of the program have been described in a separate instruction manual.
Troubleshooting
14.
Before you contact VWR Service Department
:
1.
Make sure that the operation complies with the instruction manual of the CO
2
incubator.
2.
Restart the incubator to make sure that the unit is not functioning properly. If it still does not work, disconnect the unit
again from the mains and repeat the operation after one hour.
Review the information in the table below to troubleshoot operating problems.