The same time zones are required for correct operation of computer and unit’s programs.
Alarms
10.11.
You can set parameters related to alarms.
lower alarm temp
– an alarm will be generated if the temperature drops below the value given in this field
upper alarm temp
– an alarm will be generated if the temperature rises above the value given in this field
lower alarm CO2
– an alarm will be generated if the CO
2
concentration drops below the value given in this field
upper alarm CO2
– an alarm will be generated if the CO
2
concentration rises over the value given in this field
The lower and upper alarm can only be generated after reaching the set temperature.
delay exceed alarm
– an alarm will be activated with a delay (1 min, 2 min, 5 min, 15 min) after exceeding
the permitted temperature/CO
2
concentration
delay door alarm
– a door alarm will be activated when the door is opened for the time selected by the user (5 s, 30 s,
1 min, 5 min, 10 min)
Confirms and saves changes
Cancels the entered changes
In the field "lower alarm temp" you can set a value of -1°C to - 5°C and in the field "upper alarm
temp" you can set a value of +1°C to +5°C.
Figure 43 – Alarm settings