Instruction Manual Pipettors

 VWR Standard Line
 VWR 8-channel Standard Line  
 VWR 12-channel Standard Line 

This manual is available in additional  

languages at 

www.vwr.com

 

PMS 647C

CMYK 96/54/5/27

Summary of Contents for avantor

Page 1: ...tandard Line Instruction Manual Pipettors VWR Standard Line VWR 8 channel Standard Line VWR 12 channel Standard Line This manual is available in additional languages at www vwr com PMS 647C CMYK 96 54...

Page 2: ...s are designed for microplate filling The pipet tors enable precise and simultaneous delivery of 8 or 12 preset volumes of liquid These pipettors are available in four volume ranges 0 5 10 L 5 50 L 20...

Page 3: ...ti channel manifold Contains a set of plungers and a set of flexible suspension shafts which func tion to reduce the force required to attach the tip Model Identification The volume range of the pipet...

Page 4: ...Accuracy Precision Non filter Tips L Single channel Pipettors 2 Min 0 1 40 0 12 0 10 0 2 12 0 6 0 1 2 7 1 3 Max 2 1 5 0 7 10 Min 0 5 4 0 2 8 10 1 0 2 5 1 8 5 0 1 0 0 6 Max 10 0 0 5 0 4 20 Min 2 3 0 1...

Page 5: ...o not attach the tips with a rocking movement as this may damage the shaft or plunger Observe this rule particularly with single channel pipettors of low volume range Never draw liquids directly into...

Page 6: ...In 2 and 10 L pipettors place the cap that can be purchased separately on the bottom of the pipettor shaft and slide the cap upwards to the bottom of the tip ejector M L 10000 L L 10mm 5000 L L 6mm M...

Page 7: ...ded to pre rinse the tip several times prior to use Do not pipette liquids with temperatures above 70 C NOTE When pipetting acids or corrosive solutions which emit vapors it is recommended to disassem...

Page 8: ...ISO 8655 guidelines for the maximum nominal liquid volume drawn by the pipettor and for 10 of the maximum or minimum liquid volume according to the values given Section 5 The pipettors are designed t...

Page 9: ...rm 10 aspirations and calculate the average value in mg Calculate the volume in L by multiplying the value in mg by the conversion factor Z L mg If the average aspirated volume exceeds the permissible...

Page 10: ...accuracy should not alter if the pipet ting processes including autoclaving are carried out as described in this manual If a change in accuracy occurs it is recommended to Check the calibration of the...

Page 11: ...ling elements are not sufficiently lubricated e g after repeated autoclaving procedures Incorrect aspiration Liquid with properties other than water den sity viscosity Calibrate the pipettor using the...

Page 12: ...tional product information and special offers Contact us For information or technical assistance contact your local VWR representative or visit ww vwr com 14 Warranty VWR warrants that this product wi...

Page 13: ...s Campus Northwest Business Park Ballycoolin Dublin 15 Tel 353 01 88 22 222 Email sales ie vwr com Italy VWR International S r l Via San Giusto 85 20153 Milano MI Tel 39 02 3320311 02 487791 Email inf...