LCO304
Operation | 8
B-H87.0.02.DB201-1.0
19 / 29
3. By briefly pressing the
Function key
, you can scroll through the parameters. Select
the parameter you would like to configure.
4. When you have selected the desired parameter, change the parameter to the de-
sired value with the
Up button
and the
Down button
.
5. The changes are saved after running through the entire
Configuration
menu.
STOR
appears in the display. The
Configuration
menu can be exited from any arbitrary
parameter by pressing and holding the
Function key
for 2 seconds. The changes
made up that point are saved.
Representation
Call up menu
Next parameter Change value
Save changes
Discard
changes
2s
Press: Single
step
Hold: Rapid
change
2s
2s
Outcome of an action
The
Configuration
menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
8.2.3 Configuring parameters of the configuration menu
Description
The following representation shows the available parameters and various configura-
tion options.
Prerequisite
– The
Configuration
Opening the configuration menu [
.
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the
Up button
and
Down button
.
3. The available configuration options are listed for each parameter in the following
representation.
Representation
Parameter
Values
Meaning
Input
INP
(OND
Measured variable - conductivity
RES
Measured variable - specific resistance
Temperature compensation
TCOR
OFF
Do not compensate conductivity measurement
NLF
Non-linear function for natural water in accordance
with EN 27888 (ISO 7888) Ground water, surface
water or drinking water
NACI
Compensation of weak NaCI solutions only in pure
and ultrapure water
L,N
Linear temperature compensation