10.2. Entering the buffers’ values into the meter’s memory
If the calibration with use of buffers has been chosen and the pH values set by
the manufacturer are used, there is no need to change them. However, it
should be verified whether the values correspond to those of applied buffers.
Different buffers’ values should be entered to the meter’s memory before
calibration.
To enter:
-
according to the point 9.2 choose the number of electrode (
E1
,
E2
or
E3
),
for which the points of calibration are to be changed;
-
enter the calibration mode: select the pH measurement on the
measurement screen, press and hold the
button until the
background turns red (Pic. 6A). The previous characteristic is erased;
-
press the
button, the pH measurement parameters screen will
appear (Pic. 6B);
-
press the
SOLUTION
window and choose
BUFFER
;
-
press the
button, the screen with calibration points will appear (Pic.
6C);
-
select the window with a point to be changed, a numerical keyboard will
appear to enter the value and confirm with the
button.
-
return to the measurement screen in calibration mode by double press of
the
button and calibrate the electrode in the chosen points or exit
the calibration mode by pressing the
button again;
User‘s manual for
PH-2500L
pH meter
v1.0x
Summary of Contents for 662-2380
Page 1: ...Instruction manual VWR pH METER PH 2500L EU cat no 662 2380 Version 1 1 Issue 12 2021 ...
Page 5: ... 1 I INTRODUCTION ...
Page 14: ... 10 II pH MEASUREMENT User s manual for PH 2500L pH meter v1 0x ...
Page 29: ... 25 III REDOX POTENTIAL AND TEMPERATURE MEASUREMENT ...
Page 36: ... 32 IV OPTIONS User s manual for PH 2500L pH meter v1 0x ...
Page 39: ... 35 Pic 14 ...