Instruction Manual INCU-Line Premium
Page
25
/
61
5.6.5. Priority
Can be set in terms of:
Parameters
:
The program without a ramp
– the device starts the countdown of the segment time when the set tempera-
ture is reached.
The program with a ramp
– first, the device counts down the time of the ramp and then proceeds to the seg-
ment countdown when the set temperature is reached. Regardless of whether the time of ramp elapsed.
Time:
The program without a ramp
– the device starts counting down the segment time when the program is start-
ed. Regardless of whether the temperature has been reached.
The program with a ramp
– the device first counts down the ramp time and after its expiry it proceeds to the
countdown of the segment time. Regardless of whether the temperature has been reached.
5.6.6. Protection class
The device is equipped with cartridge protection
– temperature protection, which is realized on the basis of
the temperature value measured by an independent temperature sensor, the so-called security sensor.
The main aim of the cartridge protection is to protect against uncontrolled rise or fall in temperature. At the
moment of activation, the relay disconnects the heating / cooling power supply.
Additionally, the device is equipped with
Class 3.3
– over temperature protection – when the temperature
drops below the set protection value, the power turns on automatically and after rising the temperature above
the set protection value, the
power turns off automatically.
It may happen that the device couldn’t reach the set temperature within the set time because
the reaching time was too short. In such situation the reaching time will be exceeded and the
segment’s time countdown will start when the set temperature is reached.
It may happen that the time of reaching was too short and the device failed to reach the set
temperature within the set time. Then the countdown of the segment time will start before
reaching the set temperature. Thus, the actual time of device stay in the set temperature will
be shortened.
Figure 17 - Confirmation of protection alarm class 3.3