MU 6100 H
Dissolved oxygen
ba77077e02
09/2014
43
8.2.5
Calibrating with a comparison measurement (OxiComp)
For this calibration procedure, the
Comparison meas.
setting must be set to
On
in the
Calibration
menu.
4.
Start the measurement with
<OK>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. A progress bar is displayed and
the display of the measured parameter flashes.
5.
Wait for the end of the measurement with stability control or accept the
calibration value with
<OK>
.
The calibration record is displayed and output to the interface.
6.
Using
<F1>
/
[Continue]
or
<OK>
, switch to the measured value display.
The meter displays a D.O. saturation of approx. 101.7 % in the air cali-
bration vessel.
Before calibrating via a comparison measurement, the sensor
should be calibrated in the air calibration vessel.
1.
Connect the D.O. sensor to the meter.
2.
Immerse the D.O. sensor in the reference solution.
3.
Start the calibration with
<CAL>
.
O
2
0.88
25.1
°C
OxiCal
15.02.2014
08:00
O
2
7.9
mg/l
25.1
°C
OxiComp
15.02.2014
08:00