MU 6100 H
ORP voltage
ba77077e02
09/2014
39
Freezes the mea-
sured value (HOLD
function)
With the HOLD function, you can freeze the current measured value. The dis-
played measured value stops changing until you switch the HOLD function off.
7.1.2
Measuring the temperature
For reproducible ORP measurements, it is essential to measure the tempera-
ture of the test sample.
You have the following options to measure the temperature:
Automatic measurement of the temperature by the temperature sensor
(NTC30 or Pt1000) integrated in electrode.
Measurement by an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
If you wish to measure without temperature sensor, proceed as follows:
7.2
ORP calibration
1.
Freeze the measured value with
<HOLD>
.
The [HOLD] status indicator is displayed.
2.
Release the frozen measured value again with
<HOLD>
or
<MODE>
.
The HOLD function is switched off.
The [HOLD] status display disappears.
Temperature
sensor
Resolution of the
temp. display
Temp. measurement
yes
0.1 °C
Automatic with temperature sensor
-
1 °C
Manual
1.
Measure the current temperature of the test sample.
2.
Set the temperature value with
<
><
>
.
or
In the
<OK>
/U/
Man. temperature
menu, set the temperature value with
<
><
>
.
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.