MU 6100 H
pH value
ba77077e02
09/2014
33
Continuing with two-
point calibration
7.
When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
8.
Start the measurement with
<OK>
.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
9.
Wait for the end of the measurement with stability control or accept the
calibration value with
<OK>
.
The calibration display for the setting of the nominal buffer value appears.
10. Set the nominal buffer value for the measured temperature with
<
><
>
.
11. Accept the set calibration value with
<OK>
.
The calibration display for the next buffer appears (voltage display).
12. If necessary, finish the calibration procedure as a single-point calibration
with
<MODE>
.
The calibration record is displayed.
For
single-point calibration
, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.
pH
0.0
24.8
°C
AnyCal
7.000
15.02.2014
14:15
Buffer
mV
AR
pH
7.000
24.8
°C
AnyCal
1
15.02.2014
14:15
Buffer
13. Thoroughly rinse the electrode with deionized water.