www.vtiinstruments.com
EX1200A Triggering
79
Sample Trigger Options
o
Sample Trigger Source
: sets the trigger line that moves the DMM from the
Wait-For-
Sample
state to the
Take Measurement
state.
o
Sample Trigger Slope
: indicates whether a rising or falling edge will move the DMM
from the
Wait-For-Sample
state to the
Take Measurement
state.
o
Count
: sets the number of samples to that will be taken per trigger event.
o
Delay
: indicates the delay that will occur between getting a
Sample Trigger
to taking a
measurement.
o
Interval
: when set to
Interval
, an internal time periodically sends a sample trigger. This
parameter sets the time between sample triggers.
Trigger Output Options (a trigger can be generated either when the measurement completes,
or when the read value is out of tolerance)
o
Measurement Complete Trigger
: sets the trigger line that will be pulsed when the
Measurement Complete state is reached.
o
Over Limit Trigger
: sets the trigger line to pulse when an over-limit condition occurs.
o
High Limit
: sets the high input level for the over-limit condition.
o
Low Limit
: sets the low input level for the over-limit condition.
F
IGURE
4-3:
DMM
T
RIGGER
M
ODEL
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......