4 Installation
Info
Set up simulator with two people
We recommend that assembly be carried out by two persons. Many steps will
be easier and safer this way.
4.1 Installation Location
The simulator is designed for indoor use only. Choose a setup location that com-
plies with the ambient conditions specified in
NOTICE
Possible damage to simulator due to low
temperature
Condensed moisture from the air might cause damage to the simulator.
f
If the simulator components are very cold from transporting, let them
warm up at room temperature before assembly and operation.
f
Only switch on the simulator if the simulator and room temperature are
between 10–30 °C (50–86 °F).
NOTICE
Risk of overheating due to poor ventilation
The simulator PC generates heat during operation. Overheating can cause
temporary malfunction or permanent damage to the simulator. To avoid
overheating:
f
Make sure that the operating environment is well ventilated.
f
Do not block the air vents of the simulator PC.
f
Do not allow dust to build up on the air vents.
f
Set up the simulator PC at least 10 cm (4”) away from walls or other ob-
jects which may block ventilation.
f
Do not place the simulator PC on a carpet or rug with long fibers.
f
Do not place the simulator PC in a narrow or cramped space.
f
Do not cover the simulator PC during operation.
• Do not assemble the simulator in the immediate vicinity of a radiator, heating
vent, or other heat source.
• Position the cables so that they cannot be damaged (e.g. rolling chairs, kinking)
and so that accidents (e.g. tripping) cannot occur.
• If multiple devices are used in the same space, ensure that the temperature and
humidity constraints of all devices conform to the requirements.
Eyesi Surgical Simulator – Installation and Operating Guide
18
Installation
Summary of Contents for Eyesi
Page 1: ...Platform 2 11 Eyesi Surgical Simulator Installation and Operating Guide ...
Page 74: ...Box 2 Eyesi Surgical Simulator Installation and Operating Guide 74 Appendix Packing Guide ...
Page 75: ...Box 3 75 Appendix Packing Guide Eyesi Surgical Simulator Installation and Operating Guide ...
Page 76: ...76 Notes Eyesi Surgical Simulator Installation and Operating Guide ...
Page 77: ...77 Notes Eyesi Surgical Simulator Installation and Operating Guide ...
Page 78: ...78 Notes Eyesi Surgical Simulator Installation and Operating Guide ...
Page 79: ......