BIONICARE OACTIVE  KNEE SYSTEM 

NIGHT-WRAP

OACTIVE

APPLICATION INSTRUCTIONS

CAUTION: 

Electrodes should be discarded and replaced if damaged  

or when proper adhesive tack can no longer be obtained, the gel  
has separated, or you sense a change in stimulation intensity. If in
doubt about the integrity or proper function of any electrode, replace 
it before proceeding with treatment. 

If using pain medication and/or 

other pain management devices, exercise caution when increasing
amplitude on the BioniCare device.

WARNINGS: 

This device is designed for use only with electrodes 

cleared for medical electrical stimulation.

CAUTION: 

Federal law restricts this device to sale by or on the order 

of a practitioner licensed by the law of the State in which he/she  
practices to use or order the use of the device.

WARRANTY: 

VQ OrthoCare warranties all of its products from the 

original date of purchase against defects in materials and workman-
ship. Normal wear and tear during use of a product is not considered
a defect. Contact VQ OrthoCare at 800.266.6969 for specific product 
warranty information.

© 2014 VisionQuest Industries Inc. 

VQO511659REVC

18011 Mitchell South    Irvine, CA 92614

800.266.6969       www.vqorthocare.com  

EMERGO EUROPE

Molenstraat 15

2513 BH, The Hague

The Netherlands

Phone: +31.70.345.8570

Fax: +31.70.346.7299

EC

REP

BIONICARE SYSTEM REMOVAL STEPS

STEP 1. Unhook knee wrap from straps and carefully 
peel from knee.  

Place clear plastic backing 

onto knee electrode for protection when not  
in use.
 

STEP 2. Remove brace.  

Place clear plastic 

backing onto thigh electrode.

• Do not use alcohol to clean skin surface. 
• Electrodes should only be used on dry, clean, unbroken skin surfaces.    
   Prior to Night-Wrap application wash skin with soap and water and
   completely dry.
• Do not apply skin lotion prior to application of Night-Wrap or knee wrap.
• Do not trim electrodes, as cut edges may affect the even distribution  
   of stimulation.
• Remove Night-Wrap or electrodes before showering.
• If you experience symptoms of skin irritation while using the BioniCare  
   system, discontinue use immediately and contact VQ OrthoCare’s  
   Patient Care Department at 1.800.452.7993, or consult your healthcare  
   practitioner.

 

NIGHT-WRAP AND OACTIVE LINERS

To clean Night-Wrap and liners, the electrodes should first be removed.  
The Night-Wrap and liners should be hand washed in cold water with  
mild detergent. Rinse thoroughly and air dry (do not machine dry). Never  
machine wash and/or use bleach, fabric softener, or harsh detergents.

CARE AND MAINTENANCE OF ELECTRODES AND NIGHT-WRAP

NOTE: There is no need to remove the electrodes from the Night-Wrap 
upon removal. Leave the electrodes in place and cover the electrodes 
with the clear backing to keep the electrodes clean and to prevent them 
from prematurely drying.