QUICK START GUIDE - USING MULTIPLE DC1000S WITH A LCR METER 

 

PAGE 1 OF 3 

104-174/1 

 

 

 

 

 

 

The high output impedance of the DC1000 allows it to be used with almost any manufacturer’s LCR meter.  
The unique topology also allows you to parallel up to 20 units so that up to 500A can be generated. 
The following document explains how to setup your environment to use multiple DC1000s with an LCR meter. 
In the following example we will connect 3 DC1000As together to provide 0-75Amps whilst measuring using an LCR Meter.

 

 

 

 

A) Connect RS232 + Interlocks. 

RS232. 

Use the 

77-046

 straight through 9w-9w M-F leads provided with the 

unit. 

Connect 

RS232 OUT

 from the first unit to the 

RS232 IN

 on the second 

unit. 

Connect 

RS232 OUT

 from the second unit to the 

RS232 IN

 on the 

third unit. 

 

The first unit will automatically become the “

Master

” unit and will 

control the other units in the chain. 

 

Safety Interlocks. 

Insert your interlock system or interlock override plug into the 

Interlock 

IN

 on the first unit. 

Connect 

Interlock OUT

 on the first unit to 

Interlock IN

 on the second 

unit. 

Connect 

Interlock OUT

 on the second unit to 

Interlock IN

 on the third 

unit. 

 

 

 

www.voltech.com 

Quick Start Guide 

Using Multiple DC1000s with a LCR Meter 