Vogel 65 7111 Operation Manual Download Page 1

65 7111

Electronic Digital 

Surface Roughness Measuring 

Device  

Operation manual

 

Charge battery 

Connect the charge cable to the roughness measuring device. Insert the plug in the respective socket (7) of the roughness 
measuring device. The charger then has to be connected with the mains or a computer. Charging duration is about 3 hours. 
Measurements can still be carried out during the charging procedure.  
 
 
 
 

Device care 

Protect the device against impacts and vibrations, excessive dust, moisture,  
oil stains and strong magnetic fields, etc. 

If you are not using the device, please switch it off in good time in order to keep the energy consumption as low as 
possible and charge within due time to avoid damage. 

The sensor / measuring head is the precision unit of the testing device and should be treated with special caution. After 
each test, cover the measuring head carefully with the protected area in order to avoid undesired affects. 

The enclosed standard plate for calibration should be carefully protected in order to avoid scratches that may influence 
the measuring accuracy. 

In event of malfunctions, the user should under no circumstances open the device to attempt to make repairs on it. 
Send the device back to the manufacturer along with the guarantee card and the relevant probe as well as a 
description so that it can be checked and repaired. To do so,

 

please maintain contact with our customer 

representatives or the sales department of our company. 

 
 
 
 
 
 
 

Troubleshooting:
Error Possible 

cause 

Troubleshooting 

No display

Battery is flat 

Charge the battery 

Battery symbol 
appears in the display 

Battery is flat 

Charge the battery 

Please note:

In order to exclude incorrect measurements, please perform the basic setting carefully. 
Clean the measuring instrument on a regular basis using a cotton cloth, do not use benzine or solvent.  
Keep the measuring instrument dry as humidity may cause damage to it.  
 

Electric etching or engraving may destroy the electronics, this shall void the guarantee! 

Remove the battery if the device is not used for longer periods. Do not discard of used batteries in domestic waste!  
Dispose of used batteries at designated collecting points or as special refuse. 

65 7111

Electronic Digital 

Surface Roughness Measuring 

Device  

Operation manual

C:\Users\ah\Desktop\657111_en.doc

 

 

Controls:

 

 
1. On / Off switch 
2. LC Display 
3. Function key (Ra/z) 
4. Function key (mm/inch) 
5. Protective cover 
6. Measuring head 
7. Battery charging socket 
8. Start key 
9. Reset key

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

Specification:

 

- Measuring parameter (

μ

m): 

Ra, Rz, Rq, Rt 

- Testing length (mm): 

- Sample length (mm): 

0.25 / 0.80 / 2.50 

- Assessment length (mm): 

1.25 / 4.0 

- Measuring range (

μ

m): 

Ra: 0.05 ~ 10.0 

 

Rz: 0.1 ~ 50 

- Display error: 

± 15% 

- Display deviation: 

<12 % 

- Radius of the probe tip and sensor angle: 

10 

μ

m ± 1 

μ

 

Angle: 90 +5° / -10° 

- Measuring force of the probe tips: 

”

 0.016N 

   Velocity of the measuring force: 

”

 800 N/m 

- Pressure of the sensor head: 

”

 0.5 N 

- Battery: 

3.7 V, Lithium battery 

- Dimension: 

106mm x 70mm x 24mm 

- Weight: 

200g 

- Operating temperature: 

-20°C ~ 40°C 

- Humidity: 

<90% RH 

 
 

Scope of delivery: 

 

Measuring device 

USB charge cable 

Connector 

Calibration plate (roughness standard) 

Operating instructions 

Plastic carrying case 

 
 
 

Reviews: