98 How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
Using the Vivo 65 with the FiO
2
Sensor
The FiO
2
sensor can be connected to the Vivo 65 in order to monitor
and store FiO
2
measurements. The FiO
2
sensor measures the fraction of
inspired oxygen at the air outlet of the Vivo 65. The FiO
2
measurements
will be stored in the data memory which can be downloaded to a PC and
viewed in the Vivo 65 PC software.
The FiO
2
sensor is not intended to be in contact with the patient’s body.
The FiO
2
sensor should be calibrated when first connected and then at
least once a month.
How to connect the FiO
2
Sensor
FiO
2
calibration can be performed by a clinician.
U
SAGE
T
IME
Operating temperature
50 to 104°F (10 to 40°C)
Operating pressure
700 to 1250 mbar
Response time
<12 s
Expected operating life
<6 years (in ambient air)
1 year (in 100% O
2
)
Shelf life
6 months
Note that the operating conditions for the FiO
2
sensor are different from
the Vivo 65 system conditions. If the sensor is used outside its operating
conditions the FiO
2
measurements might deviate.