57
How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
5.4 Functions and Parameters in the Vivo 65
All the parameters used for controlling the breathing by the Vivo 65 are
listed below.
Inspiratory Pressure (Insp. Pressure)
I
TEM
D
ESCRIPTION
Definition
The Inspiratory Pressure setting is used to
define the airway pressure during the inspir-
atory phase.
In the PCV-SIMV mode, this setting defines
the inspiratory pressure for the mandatory
breaths controlled by the ventilator.
Modes
PSV, PSV(TgV), PCV, PCV(TgV), PCV(A),
PCV(A+TgV), PCV-SIMV, PCV-MPV
Setting min
4 cmH
2
O (Adult), 4 cmH
2
O (Pediatric)
Setting max
50 cmH
2
O (Adult), 35 cmH
2
O (Pediatric)
Setting resolution
0.5 below 10 cmH
2
O,
1.0 above 10 cmH
2
O
Display