188 Technical Specifications
Vivo 65 users manual
Doc. 006088 En-US U-1b
Vte/MVe Sen-
sor Error
Alarm
A Vte/MVe sensor error
occurs when the sensor for
Vte/MVe measurement
fails.
Red LED, audible
alarm and a warning
message on the display.
Ambient Pres-
sure Compen-
sation Lost
Alarm
Loss of ambient pressure
compensation data or
ambient pressure sensor
failure.
Yellow LED, audible
alarm and a warning
message on the display.
Ambient Tem-
perature Com-
pensation Lost
Alarm
Loss of ambient tempera-
ture compensation data or
ambient temperature sen-
sor failure.
Yellow LED, audible
alarm and a warning
message on the display.
Humidity
Compensation
Lost Alarm
Loss of humidity compen-
sation data or humidity
sensor failure.
Yellow LED, audible
alarm and a warning
message on the display.
LED Failure
Alarm
One or more LED indica-
tors on the front panel are
broken.
Yellow LED, audible
alarm and a warning
message on the display.
Alarm for Low
Alarm Battery
Alarm battery voltage is
below alarm limit.
Yellow LED, audible
alarm and a warning
message on the display.
Alarm for
Beeper Failure
Failure of the beeper con-
trol by the treatment pro-
cessor.
Red LED, audible
alarm and a warning
message on the display.
Internal Func-
tion Failure
Alarms
Various types of internal
function failures. For defi-
nitions, refer to Vivo 65
Service Manual.
Red LED, audible
alarm and a warning
message on the display.
P
OWER
S
UPPLY
S
PECIFICATION
AC supply
100 to 240 V AC, tolerance: +10%/-20%,
50 to 60 Hz, max 300 VA.
External battery
24 V DC, tolerance: 24 V ± 6 V.
Max 7 A, 140 W.
A
LARM
S
PECIFICATION
I
NDICATION