166 Cleaning and Maintenance
Vivo 65 users manual
Doc. 006088 En-US U-1b
7.1 Cleaning the Vivo 65
Main Unit
To avoid electrical shock, disconnect the power supply to the Vivo 65
before cleaning. Do not immerse the Vivo 65 into any fluids.
• Always be careful when cleaning to ensure that you do not damage
any equipment.
• Fluid must not be allowed to enter the Vivo 65.
• Never apply any liquids directly on the Vivo 65 by spraying, splashing
or pouring. Use a moistened lint-free cloth when cleaning.
• Do not use an excessive amount of liquid when cleaning the Vivo 65.
• Do not autoclave the Vivo 65.
1
Turn off the Vivo 65 and disconnect the power supply.
2
Remove the patient circuit.
3
Disconnect all electric cables.
4
Clean the outside of the Vivo 65 using a lint-free cloth and a mild
soap solution. If the Vivo 65 needs to be surface disinfected, this can be
done using:
– Ethanol 70%
– Isopropyl alcohol
– Ethanol 90%, Methanol 9.5%, Pyridine 0.5%
5
Reconnect the patient circuit. Make sure all parts are dry before the
Vivo 65 is put into operation.