4 Operating the SuperPump
4.1
Powering On the System
Turn on the power switch on the back of the Controller. The screen on the front of the Controller
will light up and indicate that it is Initializing. It will then automatically begin a Homing Pump
sequence.
V 3.0.5 2014/02/20
0.0 % Gain
0.0 ml/Stroke
Initializing
After automatically homing the Pump, the screen will appear as below. The most recently used
waveform source will appear in the Current Waveform section.
If the Pump is not connected to the Controller when the system is powered on, Pump Error will
be displayed. Power off the Controller, connect the Pump cable, and power the Controller on to
clear the error.
Firmware Version
Cycle Counter
Current
Waveform
Sync Output
Delay
Current
Amplitude
Current Stroke
Volume
Firmware Date
Menu
Dial
Amplitude
Dial