![Vitrek Xitron 6000-2 User Manual Download Page 99](http://html1.mh-extra.com/html/vitrek/xitron-6000-2/xitron-6000-2_user-manual_1037510099.webp)
Appendix B
99
Mode Effect:
Add (0.04 + 5F/F
C
)° for voltage: current phase.
Ranging Effect:
Add (0.02 + 3F/F
C
)° per range difference between inputs.
Fundamental Frequency Measurement
Accuracy:
0.01% + (0.03% divided by measurement period in seconds).
Min. Fundamental:
5% of range.
Frequency Range:
0.0005Hz to 200kHz.
Technique:
Period measurement of filtered signal using a “tracking” bandpass
filter having a bandwidth of 0.8 to 1.2 times the measured
fundamental frequency.
Harmonic Rejection:
Detects fundamental component having an amplitude down to 30%
of total signal.
Measurement Filter:
Optional “moving average” filtering may be applied.
Response Time:
<1.5 measurement periods.
Harmonic Analysis
Normal Analysis:
User definable window period between 1 and 512 fundamental
cycle periods, using rectangular window with no overlap and with
no gaps for fundamental frequencies <100Hz. Automatically sized
(256 to 4096 point) Fast Fourier Transform with average sampling
frequency maintained within 10ppm of relevant ratio from the
fundamental frequency.
Windowed Analysis:
User definable window period between 8 and 512 fundamental
cycle periods, using a modified Blackman-Harris window with no
overlap and with no gaps for fundamental frequencies <100Hz.
Automatically sized (256 to 4096 point) Fast Fourier Transform
with sampling frequency maintained within 10ppm of relevant ratio
from the fundamental frequency.
Frequency Range:
0.0005Hz to 150kHz, up to 2047th harmonic.
Abs. Single Harmonic:
(0.05 + 2F/F
C
)% of r 0.01% of range + (2F/F
C
)% of total
signal. F is the frequency of the harmonic.
Abs. Harmonic Range:
(0.05 + 2F/F
C
) % of r (0.01% of range + (2F/F
C
) % of total
signal).
√
(number harmonics). F is the frequency of the
fundamental.
THD Accuracy:
(0.01 + 2F/F
C
) %. (range / total signal level) .
√
(number
harmonics). F is the frequency of the fundamental.