background image

COPD Screener

copd-6™

Guide for Healthcare  

Professionals

Summary of Contents for copd-6

Page 1: ...COPD Screener copd 6 Guide for Healthcare Professionals...

Page 2: ...Prepare Device 2 Set test subject s age Press B or C to change age Press E to set 1 D Switch on 3 Set test subject s height in inches Press B or C to change height Press E to set...

Page 3: ...C to change gender Press E to set 5 Set test subject s population group Press B or C to change population group Press E to set Letter Population Group Ref c Caucasian NHANES III AA African American NH...

Page 4: ...nstrate to test subject v Fully breathe in hold v Insert and lightly bite the mouthpiece v Then blow as hard as possible for 6 seconds 1 Insert disposable SafeTway mouthpiece 4 Press E to show best te...

Page 5: ...dex Low v Refer for spirometry v COPD not likely Possible COPD Arrows always show best test Positive v Refer for spirometry v COPD classification I to IV Test Result Normal Arrows always show best tes...

Page 6: ...ays numbers for each blow until E pressed 3 D for 3 seconds to switch off Cleaning and disinfecting The outer surfaces should be cleaned every week more often if necessary The use of a disposable disi...

Page 7: ...III Stage IV 0 1 2 3 FEV FEV Ratio and FEV Predicted FEV FEV 0 7 FEV FEV 0 70 FEV 80 Pred FEV FEV 0 70 FEV 80 Pred FEV FEV 0 70 FEV 50 Pred FEV FEV 0 70 FEV 30 Pred FEV Predicted 80 80 50 30 Obstruct...

Page 8: ...ger Stra e 64a 20257 Hamburg Germany Tel 040 547391 0 Fax 040 547391 40 e mail info vitalograph de www vitalograph de Vitalograph Inc 13310 West 99th Street Lenexa Kansas 66215 USA Toll Free 800 255 6...