vita Vacumat 500 Operating Manual Download Page 1

VITA - Vacumat 500

Operating Manual

Summary of Contents for Vacumat 500

Page 1: ...VITA Vacumat 500 Operating Manual...

Page 2: ...ption Vacuum Block E 5 4 6 Description Slow Fast cooling Block F 6 4 7 Description of Keyboard Lift position keys Block G 6 4 8 Operating Block B Pre drying temperature Pre drying time 7 4 9 Operating...

Page 3: ...Volts A C 50 60 Hz max 0 2 KW Vacuum pump optional Type PJ 9080 023 3 110 230 Volts A C 50 60 Hz IP 20 Weight approx 6 4 kg Supply schedule Special shipping carton containing 1 VITA VACUMAT 500 furna...

Page 4: ...e the surface of the insulation at the lower edge of the firing chamber regularly with a damp cloth This applies also to the O ring on the lift support plate The casing can likewise be cleaned using a...

Page 5: ...gram 3 flashes when PRE DRYING is active Display shows 1 Stand by temperature 200 C 800 C 2 PRE DRYING temperature 200 C 800 C 3 PRE DRYING time 0 00 min 40 00min Acivates Input Mode 1 Changes PRE DRY...

Page 6: ...ime of END temperature is active Display schows 1 END temperature 200 C 1200 C 2 HOLD time of END temperature 0 00 min 40 00 Min Actives Input Mode 1 Changes END temperature 2 Changes HOLD time for EN...

Page 7: ...ashes when COOLING ist active Display shows COOLING temperature 300 C 1000 C Activates Input Mode 1 COOLING temperature 300 C 1000 C 2 Select programm FAST COOLING of firing chamber to 50 C below STAN...

Page 8: ...Valid Input 200 C 800 C Change PRE DRYING temperature NO YES enter value using keyboard and press Enter Key PRE DRYING LED ON LCD shows PRE DRYING temperature Valid Input 0 min 40 00 min E N D Press E...

Page 9: ...lue using keyboard and press Enter Key HEATING UP LED ON LCD shows value C min E N D Press Enter Key Input Mode may be interrupted at any time by pressing the STOP Key Invalid entries will result in e...

Page 10: ...er value using keyboard and press Enter Key END temperature LED ON LCD shows END temperature C Valid Input 200 C 1200 C E N D Input Mode may be interrupted at any time by pressing the STOP key Invalid...

Page 11: ...ja NO flashing VACUUM LED LCD shows VACUUM time in min sec see flashing VACUUM LED LCD shows VACUUM time in min sec Valid Input 0 min 40 00 min VACUUM time Based on the selected temperature rising ra...

Page 12: ...nd press Enter Key COOLING LED ON LCD shows COOLING temp C Press Enter Key Input Mode may be interrupted at any time by pressing the STOP key Invalid entries will result in error message Err Activate...

Page 13: ...icates the value in millivolt After two minutes in the hold phase at a temperature of 1000 C a short signal sounds which indicates the lift keys UP DOWN may now be used for adjustment Using the UP DOW...

Page 14: ...284 start firing program when temperature of firing chamber is higher than stand by temperature Selection 1 start when temp is higher than stand by temperature Selection 2 start at stand by temperatu...

Page 15: ...screw the six screws on the to casing of the furnace 3 Lift off top casing of the furnace 4 Unscrew and remove tthe screws 22254 on the top cover of ring chamber 5 Disconnect all wires from the thermo...

Page 16: ...e factory stored value is 15 seconds The time selected must be longer than the time the lift drie takes to move from the lower to the upper position If the time is too short the motor will switch off...

Page 17: ...ufacturer s instructions VITA OMEGA 900 Progr no Pre Drying C min C min Temp approx C min VAC Oxidation firing 41 Follow manufacturer s instructions 1st opaque firing powder 42 600 2 00 75 900 2 00 x...

Page 18: ...elf The vacuum time indicated in the VAC column by an X resuluts automatically from the given temperature rise C min The vacuum switches on at the biginning of the temperature rise and switches off ag...

Page 19: ...g of the temperature rise and switches off again when the end temperature is reached Additional programs Progr no Pre Drying C min C min Temp approx C min VAC Metall Corrector 91 600 2 00 75 1 040 1 0...

Reviews: