Maintenance
Routine Maintenance
: ali-Q 2 will require minimal, if any, maintenance, if used as intended. VistaLab recommends
performing routine maintenance and calibration verification at least once a year. Routine maintenance includes:
Cleaning and disinfecting outer body, nozzle cover, nozzle insert and filter
Replacing the membrane filter
Replacing nozzle insert (as-needed)
Aliquot mode calibration verification
Cleaning / Decontaminating
:
-
Chemical - Occasionally wipe outer housing of the ali-Q 2 and the nozzle cover with a damp
cloth of soapy water or diluted alcohol solutions (e.g. 70% isopropanol).
WARNING!
ALWAYS use protective gloves and eye protection. Avoid the risk of contamination!
-
UV Sanitizing - The complete ali-Q 2 unit (outer housing) can withstand typical output (254 nm
wavelength) from UV sterilization lamps.
NOTE: The effects of the UV exposure will likely cause discoloration over time.
Autoclaving:
NEVER AUTOCLAVE ALI-Q 2!!
Only the nozzle cover, nozzle insert and membrane filter can be
detached from the unit
and separately autoclaved at:
•
121
°
C (250
°
F); 30 psi (2 bar absolute) with a holding time of at least 15 minutes.
Replacing the Membrane Filter
:
1
Unscrew the purple nozzle cover
.
2
Remove the membrane filter.
3
Place new membrane filter with the thicker end facing downwards into the silicone nozzle
insert and thinner end into ali-Q 2.
4
Screw the nozzle cover assembly back onto the ali-Q 2.
Replacing the Nozzle Insert
:
1
Unscrew the purple nozzle cover
.
2
Remove the membrane filter and silicone nozzle insert from nozzle cover.
3
Insert new silicone nozzle insert into nozzle cover. Place membrane filter with the thicker end
inserted into the silicone nozzle insert and thinner end into ali-Q 2.
4
Screw the nozzle cover assembly back into the ali-Q 2.