vicosys
®
6300
Commissioning
ADVICE
Factory setting: The device searches for a DHCP server during initialization in the LAN. If
no DHCP server is available, a static IP address is used (default: 192.168.3.180).
If the device cannot be found, a monitor can be connected and the IP can be read off
during initialization.
The settings can be changed via vcwin.
ADVICE
The search function in the TCP/IP tab of the dialogue Communication Parameters only
detects devices in the same subnet. For devices in different subnets, the configuration
data must be entered manually.
6.4.2 Create a serial connection
The Serial interface on the device is used for serial communication.
1. Connect the device directly to the PC using a RS232 cable.
2. Start the operating software vcwin on the PC.
Configure the interface by calling
Communication > Interface
in the menu.
1
2
4
5
3
3. Switch to the RS232 tab.
4. Select the COM interface of your computer.
5. Set the baud rate to 9600. Once connected, the baud rate can be adjusted.
6. Deactivate RTS/CTS (Hardware handshake of the RS232 interface).
7. Click the [OK] button to accept the values.
36
999.995.428.10-en-2.1
© Vision & Control GmbH 2022