Operating instructions
EN
Rev. 9/6480-2021-09
Read these instructions carefully
before using the thermometer
CAUTION: read the warnings
PRO
06480
Page 1: ...Operating instructions EN Rev 9 6480 2021 09 Read these instructions carefully before using the thermometer CAUTION read the warnings PRO 06480...
Page 2: ...ngs modes 9 Protective cap 10 Label with serial number 11 Battery door 4 x AAA with brief instructions 12 Special lanyard with additional brief instructions to be worn around the neck 10 11 12 9 6 5 8...
Page 3: ...nd allows the rapid sta bilization of the thermometer at room temperature the display shows a countdown indicating the time required for its stabilization To make a measurement it is necessary to wait...
Page 4: ...mperature than the room where the thermometer is used in these cases wait for the stabilization of the subject s temperature or in case it is not pos sible take the measurement on the closed eyelid pa...
Page 5: ...e the thermometer in water or other liquids and keep it away from sources of heat and out of direct sunlight If water seeps into the thermometer contact your Dealer im mediately for Technical Service...
Page 6: ...emperature varies according to the measurement s site and throughout the day also in response to physical or mental effort for example a baby s crying Moreover the body temperature can be affected by...
Page 7: ...h twice slowly and then the display will show the acquired temperature VisioFocus PRO is now ready to take a measurement For more details about MQCS see par 8 To take a reading press the FACE button a...
Page 8: ...ou are taking the reading the lights are harmless Precision is not guaranteed but such reading can be considered a valid approximation of one s body temperature Such reading is also indicated when one...
Page 9: ...33 8 176 0 F temperature range Proceed as you would for a forehead reading but press the home button the display will light up in green and will show the home symbol NOTE this intended use is not subj...
Page 10: ...g in the room to minimize the cooling effect on the subject patient If necessary these settings can be modified as follows 1 while the thermometer is off or in stand by mode press the Mem button and h...
Page 11: ...ferent temperature and you cannot wait too long to have the thermometer stabilized to room temperature for example if you move from room to room in a hospital department especially if rooms have diffe...
Page 12: ...usly handled At this point you have 2 options 1 make the MQCS as described in par 8 1 or 2 without touching the thermometer wait until the countdown has run its course and automatic quick calibration...
Page 13: ...ggested by international guidelines DO NOT USE the thermometer for at least 30 minutes after cleaning When this thermometer is intensively used in hospitals or similar environments a pe riodic calibra...
Page 14: ...CRIPTION when you press the face button the projection display shows Hi 4 and the value alternately PROBLEM the room temperature is between 40 1 and 45 0 C 104 1 and 113 1 F SOLUTION the temperature r...
Page 15: ...it for unit operation 80 0 C 176 0 F SOLUTION reading with face button please make certain that you ve pressed the correct button with regard to the measured object and that the warnings have been met...
Page 16: ...check that the waveguide fig 1 is not soiled or damaged if it is clean it as indi cated in par 10 or contact your Dealer for technical service check that the thermometer is perpendicular to the forehe...
Page 17: ...yed alternately 2 ASTM E1965 98 2016 laboratory accuracy requirements in the display range of 37 to 39 C 98 to 102 F for IR thermometers is 0 2 C 0 4 F whereas for mercury in glass and electronic ther...
Page 18: ...ses in accordance with an EN ISO 9001 2015 and EN ISO 13485 2016 certified Quality System with GMP requirements and with the following standards EN 60601 1 EN 60601 1 2 EN 60601 1 6 EN 60601 1 11 EN 6...
Page 19: ...mproper use The warranty is also voided if the product is tampered with damaged or used improperly the label on the back bearing the serial number is removed damaged or rende red illegible the product...
Page 20: ...66 US 7 651 266B2 US 8 128 280 US 8 821 010 EP 1 283 983 EP 1 886 106 EP 2577242 B1 KR 10 1898897 CN 103026192B Other international patents pending Manufactured in Italy by TECNIMED srl P le Cocchi 12...