<5. Startup>
5-3
IM 11Y01D01-01EN
NOTE
If you are using the Field Communicator 475 configuration tool, the following screen may appear
depending on the factory default settings of the TDLS8000. This is a notification that the same
measured value was read multiple times within the same analysis period. Since this is not
a problem with the TDLS8000 operation, select “YES.” This screen will appear every time a
connection is made. The analysis period is a fixed adjustment value assigned to each TDLS8000
and cannot be changed.
5.2
Connecting to the YH8000
This section describes how to connect the YH8000 HMI unit to the TDLS8000 and provides a
basic description of the related screens. For a detailed explanation of the YH8000, see “8.
5.2.1
Initialization and Connection Procedure
If you are connecting the TDLS8000 and YH8000 in a one-to-one configuration, the factory
default settings can be used. The settings need to be changed when you are connecting the
TDLS8000 to an existing LAN network or when you are connecting multiple TDLS8000s or
YH8000s.
The initial IP address settings are shown in Figure 5.1.
Table 5.1
Factory default IP address settings
Model
IP address
Subnet mask
Default gateway
TDLS8000
192.168.1.10
255.255.255.0
192.168.1.254
YH8000
192.168.1.100
255.255.255.0
192.168.1.254
To change the TDLS8000 IP address and subnet mask from the YH8000, follow the procedure
below.
(1) Connect the YH8000 using the factory default IP settings.
(2) Change the TDLS8000 IP address and subnet mask (see section 5.2.2).
(3) To change the IP address of multiple TDLS8000s, connect to the TDLS8000s one at a time
and change the IP address and subnet mask.
(4) If necessary, change the YH8000 IP address and subnet mask (see section 5.2.2).
(5) Change the YH8000 destination IP address (see section 5.2.3).
The procedure from turning on the YH8000 to connecting it in step (1) above is provided
below. The procedure assumes that the TDLS8000 has the factory default IP address settings.
To connect to a TDLS8000 whose IP address has been changed or to connect to multiple
TDLS8000s, see “5.2.3 Connecting to the TDLS8000”.
(1) Complete the wiring of the TDLS8000 and YH8000 according to the instructions in “3.2.8
Connecting an Ethernet Cable”.
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...