<9. Inspection and Maintenance>
9-23
IM 11Y01D01-01EN
9.4.2
Configuration
Offline validation configuration menu:
[HART] “Detailed setup>>Validation>>Offline validation #”
[YH8000]
“
>>Configuration>Validation>>Offline Validation #”
The setup parameters required to manually execute offline validation are indicated for each of the
above submenus (tabs on the YH8000). Here, offline validation 1 will be used as an example.
l
Parameter
Parameter name
(HART)
Description
Offval1 gas type
Selects the type of validation 1 check gas (two-gas measurement only)
Offval1 gas conc
Enters the concentration of the validation 1 check gas
Offval1 pres mode
Selects the pressure mode for validation 1 execution (*1)
Offval1 pres fix val
Enters the pressure for when Offval1 pres mode is set to Fixed
Offval1 temp mode Selects the temperature mode for validation 1 execution (*1)
Offval1 temp fix val
Enters the temperature for when Offval1 temp mode is set to Fixed
Offval1 OPL mode
Selects the optical path length mode for validation 1 execution (*1)
Offval1 OPL fix val
Enters the process optical path length for when Offval1 OPL mode is set to Fixed
*1:
Process parameter: Uses the process parameter value
Fixed value: Set to a fixed value
l
Valve
Parameter name
(HART)
Description
Offval1 auto vlv man
Selects whether to enable automatic valve control through the SV terminal during
manual validation 1 execution.
Offval1 gas purg time For automatic execution. Set these when performing automatic execution (see section
9.8.2).
Offval1 prc purg time
l
Auto time
For automatic execution. Set this when performing automatic execution (see “9.8.2
9.4.3
Execution
Before starting offline validation, check that the piping and offline validation settings are correct.
Here, offline validation 1 will be used as an example.
Execution menu path:
[HART] “Diagnosis/Service>>Validation>>Manual>>Manual offline val 1”
[YH8000]
“
>>Execution>>Validation>>Manual>>Offline Validation 1”
(1) Starting an offline validation
On HART or YH8000, open the above menu, and start offline validation. If automatic
valve control is enabled, a message stating that the valves will be automatically controlled
will appear.*1 In this case, you do not need to manually control the valves during offline
validation.
(2) Purging with check gas
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...