<6. Configuration>
6-20
IM 11Y01D01-01EN
Item name
(HART)
Initialization target
Setting data
All parameters settable from YH8000 and HART
(All parameters in section 6.11.2 except for “User info.”)
AI/AO cal data Input/output calibration data of AI/AO
Calibration data
Zero/span calibration data
User info
TDLS8000 tag, IP settings, HMI (YH8000) user password, HART address,
HART 8-character short tag, write protect password
NOTE
When initialization is executed, the TDLS8000 restarts automatically.
6.11.2
Parameter Initial Value List
The initial values of parameters at the time of shipment are as follows.
Parameters with ○ in the “User specification” column are initialized to the values specified by the
customer at the time of ordering if the values were specified by the customer at that time.
l
Process parameters
Parameter
Initial value
User
specification
Min. – Max.
OPL (Optical Path Length)
0.660[m]
○
0.01 – 100[m]
Pressure mode
Active input
Select in the screen
Pressure value for fixed mode
101.325[kPa]
0.1 – 10,000[kPa]
Pressure active input source
AI-1
Select in the screen
Pressure value at 4mA
40[kPa]
0 – 10,000[kPa]
Pressure value at 20mA
200[kPa]
0 – 10,000[kPa]
Pressure backup mode when AI-1 input is out of range or under AI
calibration
Backup value
Select in the screen
Pressure backup set value
101.325[kPa]
0.1 – 10,000[kPa]
Temperature mode
Active input
Select in the screen
Temperature value for fixed mode
25[deg C]
-273 – 3,000[deg C]
Temperature active input source
AI-2
Select in the screen
Temperature value at 4mA input
0[deg C]
-273 – 3,000[deg C]
Temperature value at 20mA input
100[deg C]
-273 – 3,000[deg C]
Temperature backup mode when AI-2 input is out of range or
under AI calibration
Backup value
Select in the screen
Temperature backup set value
25[deg C]
-273 – 3,000[deg C]
Temperature offset for active ambient method
-6[deg C]
-100 – 100[deg C]
l
Non-process parameters
Parameter
Initial value
Min. – Max.
Non-process OPL
0[m]
0 – 10[m]
Non-process pressure
101.325[kPa]
0.1 – 10,000[kPa]
Non-process temperature mode
Fixed
Select in the screen
Non-process temperature
25[deg C]
-273 – 3,000
[deg C]
Non-process temperature compensation coefficient for active
ambient method
0.5
0 – 1
Non-process concentration for measurement gas 1
0[ppm]
0 – 1E6[ppm]
Non-process concentration for measurement gas 2
0[ppm]
0 – 1E6[ppm]
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...