<6. Configuration>
6-14
IM 11Y01D01-01EN
Measurement absorption spectrum
Measurement received optical signal
Reference cell absorption spectrum
Reference cell received optical signal
6.9.5
Communication Address Setting
Configure the address settings for TDLS8000 and HART communication.
l
IP address setting of TDLS8000
Setup menu path:
[HART] “Detailed setup>>System>>Communication>>TCP/IP>>Set IP settings”
[YH8000] “Configuration>>System>>Communication>>TCP/IP”
NOTE
When the IP address is changed, the TDLS8000 restarts automatically.
NOTE
When the IP address of the TDLS8000 is changed via the YH8000, the YH8000 connection
settings need to be configured again. For details, see “5.2.3 Connecting to the TDLS8000”.
l
HART address setting
Setup menu path:
[HART] “Detailed setup>>System>>Communication>>HART output”
[YH8000] “Configuration>>System>>Communication>>HART”
“Loop current mode” is a setting related to multi-drop mode. For details on multi-drop mode, see
6.9.6
Moving Average Count for Analysis Values
Configure the setting for how many analysis cycles of spectrum data to take the moving average.
Increasing this value enables a more stable analysis result to be obtained but the response time
will increase. For explanations on the analysis cycle and average count, see “Appendix 5 What
3rd Edition: Jun. 10, 2016-00
Summary of Contents for TDLS8000
Page 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Page 25: ...Blank Page...
Page 45: ...Blank Page...
Page 115: ...Blank Page...
Page 151: ...Blank Page...
Page 231: ...Blank Page...
Page 239: ...Blank Page...
Page 243: ...Blank Page...
Page 261: ...Blank Page...
Page 303: ......
Page 305: ......
Page 307: ......
Page 309: ......
Page 311: ......
Page 313: ...Blank Page...