background image

INSULATION TESTER INSTUCTION MANUAL

1. Introduction

The digital megohmmeter uses the DC-excited transducer to convert DC 12V to DC high voltage 1000V/2500V. The main  
amplifier tests the insulation voltage by the current ratio measurement of traditional megger. The input terminal uses micro-
current   to   test   anti-interference   circuit   and   the   output   terminal   adopts   the   division   function   of   double   integrating   digital 
voltmeter to process the data conversion. It is characteristic of strong loading capacity, high anti-interference of electric field, 
portable operation, wide range, backlight display and so on. It could also be used to test commercial electric power. With its  
stable performance and high grade appearance, it is widely used to test the insulated voltage resistance for electrical equipment, 
instrumentation, cables and so on.

2. Appearance Description

1.2.3.4. Voltage Selection Switch (It includes 

AC750V/200M/ 2GΩ/20GΩ

)

5. Voltage Selection Switch (

1000V/2500V

) (except AC750V)

6. Power Switch: Self-locking power switch (POWER).
7. High voltage indication: LED display.
8. Test Switch
9. LCD display: display the measuring value and unit symbol.
10. Instrument mode
11. L: jack for circuit under test.
12. G: Protection jack. When the object under test is requested to add 

the protection ring to eliminate leakage effects, the electrode wire 
of protection ring should be connected to "G" jack.

13. ACV: Input end for AC voltage measurement.
14. E: Jack for object under test.
15. Power adapter jack 

).

3. Technical Features

3.1. General Features
(1) Display: 84.8×59.8mm LCD display. The max display “1999”.
(2) Over range indication: the MSD shows "1" when it is over ranged.
(3) Alarm function: the meter will alarm automatically when the tested resistance is below the lower range and the reading is 

invalid. 

(4) Power supply: #5 batteries LR6 (1.5) ×8.
(5) Low battery indication:

 “

displays 

(6) Operation environment: temperature  0

40

30% R.H- 85% R.H 

(7) Weight: Approx. 630g (including batteries)
3.2 Technical index

Basic function

Range

Basic accuracy

Output Voltage

1000V/2500V

±10%

Short-circuit current

<4mA

Insulated Resistance 1000V

200MΩ:6.0-199.9MΩ

±(5%±5 digit

2GΩ:0.06-1.999GΩ

±(5%±5 digit

20GΩ:0.6-19.99GΩ

±(5%±5 digit

Insulated Resistance 2500V

200MΩ:5-199.9MΩ

±(5%±5 digit

2GΩ:0.05-1.999GΩ

±(5%±5 digit

20GΩ:0.5-19.99GΩ

±(5%±5 digit

200GΩ:5.0-199.9GΩ

±(5%±5 digit

Voltage Measurement

AC750V

±

1% +6 digit

Jack Position

Insulated Resistance: L

E  AC750V: ACV  G

ACV750V input impedance: 1MΩ.

15