Victor VC60B+ Instruction Manual Download Page 1

 

- 1 - 

INSULATION TESTER 

INSTUCTION MANUAL 

一、

GENERAL

 

This instrument adopts the DC Voltage convertor of low 

consumption  and  high  converter  ratio  inductance  energy 
storage, to change the 9V Voltage to 250V/500V/1000V DC 
Voltage.  It  uses  digital  bridge  to  make  the  insulation 
resistance  measurement  had  the  features  of  user-friendly, 
high  extent  range,  backlight  display,  data  hold  and  auto 
power off. The whole instrument is stylish and superior grade 
in  appearance  and  high  stable  in  performance,  it  can  be 
operated by  both hands  through using a shoulder  strap. It  is 
most  suitable  for  the  needs  of  insulation  resistance  check, 
such as electric, cable, electrical and mechanical equipment, 
telecommunication equipment and electrical facilities. 

二、

FRONT PANEL DESCRIPTION 

 

 

 

 

 

 

 

1

2

3

4

Voltage selection switch

AC750V/500V/250V/1000V

5

Resistance Range selection switch

RANGE

6

Power switch

auto-lock power switch

POWER

7

High voltage indicator

LED display.   

8

Measurement button. 

9

LCD display

Display the measured data and units

。 

10、

Instrument model. 

11

L: 

Terminal for connecting the measured circuit. 

12

G

Terminal for protection, connect the electrode wire of 

protection  loop  to  “G”  terminal  when  the  measured 
object is required to add the protection loop to eliminate 
the leakage effect. 

13

ACV:

 Input terminal of AC Voltage measurement. 

14

E

Terminal  for  connecting  the  ground of  the  measured 

object. 

15

Terminal of AC adaptor

 

三、

TECHNIC PROPERTY 

1

General Property 

1

Display

84.8

×

59.8mm LCD display, max.

1999

”。

 

2

Over range indicator

display

1

”。

 

3

Power

5#  battery  LR6  (1.5V) 

×

6  (or  external  AC 

adaptor),  no  battery  indicator

Auto  power  off 

(approx. 15miutes after turning on) 

4

Power consumption

consumption is less than 300mw 

at unload measurement. 

5

Operation  environment

Temperature  0

°C

40

°C

Relative humidity 30%RH

85RH. 

6

Dimension

175(L) 

×

110(W) 

×

70(D) mm

 

7

Weight

630 g (including batteries). 

2

Technical property 

Measured 

voltage 

250V

±

10% 

500V

±

10% 

1000V

±

10% 

RANGE 

 

0.1M

Ω—

20

M

Ω

 

0.1M

Ω—

50

M

Ω

 

0.1M

Ω—

100

M

Ω

 

 

20M

Ω—

500

M

Ω

 

50M

Ω—

1000

M

Ω

 

100M

Ω—

2000

M

Ω

 

Accuracy 

±(4% of reading ± 2 digits)

 

Shot current 

1.8mA 

1.8mA 

1.6mA 

Median 

resistance 

2 M

Ω

 

2M

Ω

 

5M

Ω

 

Terminal 

position 

L    .    E 

ACV750V    Accuracy

:±

(1.0% of r 6 digits). 

Terminal position

ACV    G 

Input impedance

  1M

Ω。

 

Frequency response

50

200

Hz 

Note

Median  resistance

——

to  ensure  that  the  voltage  of 

both  sides  of  measurement  is  not  less  than  90%  of  lower 
limit  of  resistance  measurement  of  measurement  voltage 
normal value   

四、

OPERATION DESCRIPTION

 

1

Open  the  battery  case  and  set  up  6pcs  of  5#  batteries 
(see  the  following  pictures),  note  the  polarity  of 
batteries. 

 

 

 

 

 

 

 

 

2

Press the button of

POWER

3

Select the right voltage (250V/500V/1000V/AC750V) according 

to the requirement of measurement. 

4

Select the right range (RANGE) (except AC750V), according to 

the requirement of measurement. 

5

Connect  the  electrode  of  the  measured  object  to  the  relative 

terminal of the instrument. 

6

When  measuring  cable,  connect  G  Terminal  to  the  protection 

loop. 

7

Press  the  button  of  measurement  and  start  it,  turn  the 
knob right to lock the button; when the value is stable, 
get the reading from LCD.

 

8

Connect  the  input  line  “E”  to  the  ground  of  the 

measured  object,  and  “L”  to  the  circuit  of  the 
measured object; require line “L” to hang in the air.

 

9

2000M

Ω。

When “1” is displayed, over range is indicated, 

and  should  turn  to  the  higher  range  to  get  reading; 
When  the  range  button  is  in  “ 

  “position,  it 

expresses  that  the  insulated  resistance  exceeds 
2000MΩ.

 

10、

Hang the instrument on your neck to save your hands 

15