IND110 Technical Manual
5-8
Calibration
The above chart illustrates the calibration and setup functions. Position the DIP switch
located on the front panel to select the desired function. After calibration is complete,
set all switches to OFF to store the settings.
SWITCHES LED
status
OPERATING
MODE
1 2 3 4 5 6 LO HI
CONDITIONS NOTES
on on Weight
below
setpoint
off on Weight
between
LO
& HI setpoint
off off Weight
above
HI
setpoint
Operational
status
Normal
conditions
Normal
Weigh
Mode
Off
Off
Off
Off
Off
Off
flash flash
Error
Analog=0 mA.
Setpoint=OFF
off flash
Coarse
zero
adjustment mode
Zero
calibration
On
Off
Off
Off
Off
Off
flash off Fine
zero
adjustment mode
off flash
Coarse
span
adjustment mode
Span
Calibration
Off
On
Off
Off
Off
Off
flash off Fine
span
adjustment mode
off flash
Coarse
HI
setpoint
adj. mode
HI setpoint
setting
Off
Off
On
Off
Off
Off
flash off
Fine HI setpoint adj.
Mode
off
flash
Coarse LO setpoint
adj. Mode
LO setpoint
setting
Off
Off
Off
On
Off
Off
flash off
Fine LO setpoint
adj. Mode
Use
4
key to
toggle from
coarse to fine
modes.
LO LED flashes
for coarse.
HI LED flashes
for fine.
Use + key to
increment or
Use – key to
decrement
analog output
voltage.
Press and hold
key for faster
speed
adjustment.
METTLER TOLEDO
IND110
PWR
Lo
Hi
SET key
INC key
DEC key
-
+
4
Setpoint
Output
4/20 mA
Output
DC Power
Input
Power ON LED
Hi Setpoint LED
Calibration switches
Lo Setpoint LED
Analog load
cell input
Summary of Contents for IND110
Page 1: ...IND110 Load Cell Signal Converter Technical Manual 17041600A 5 04 00...
Page 2: ......
Page 7: ...For your notes...
Page 10: ...IND110 Technical Manual 1 2 For your notes...
Page 14: ...IND110 Technical Manual 3 2 For your notes...
Page 26: ......
Page 28: ...6 2 For your notes...
Page 29: ......