VICI DM6804A + Instruction Manual Download Page 1

THERMOMETER

C/ F

TYPE-K/E/J/T/B/N/R/S

MAX

MIN

8

1

2

3

6

7

5

4

9

INTELLIGENT DIGITAL THERMOMETER 

INSTRUCTION MANUAL 

 

1.

 

GENERAL DESCRIPTION 

This multifunctional 4 ½ digital thermometer with highly stable performance and high accuracy is driven by 
9V battery. With following features, it’s an ideal intelligent tool for temperature measurement:   

1.1

 

Support K, E, J, T, N, B, R, S thermocouples; 

1.2

 

 and 

 display, data hold; 

1.3

 

Dual display: display the Max/ Min and relative mV value of input temperature; 

1.4

 

Dual channel: measure the T1, T2 and T1-T2 value.. 

1.5

 

Memory function: manually record 2500pcs temperature points, and automatically record 6000pcs 
temperature points. Recording interval can be adjusted. The longest recording time is about 99 hours. 
 

2.

 

SAFETY REGULATION FOR OPERATION 

This instruction manual covers information on operation and safety. For safety and better use, please read the 
relevant information carefully before your operation! 
2.1

 

Do not operate this instrument in occasions with stem or inflammable and explosive gas. It’s very dangerous.   

2.2

 

Do not input voltage higher than 30V for both terminals. Do not connect T1 and T2 in parallel. 

2.3

 

Wipe the case with a damp cloth and detergent. Do not use abrasives and alcohol to clean the meter. 

2.4

 

Do not disassemble the meter without professional direction. 

2.5

 

For accuracy, please wait 5 minutes to warm up the meter after power on 

2.6

 

Please contact the factory if the meter’s accuracy has big error. 

 

3.

 

GENERAL SPECIFICATIONS 

3.1

 

Display: 4 ½ digit large LCD, Max display 19999. 

3.2

 

Sampling rate: 2.5 times/sec. 

3.3

 

Over range display: “

OL

”. 

3.4

 

Low battery indication: LCD displays “

”. 

3.5

 

Working conditions: -10

50

, relative humidity<80%. 

3.6

 

Storage conditions: -20~60

, relative humidity<80%. 

3.7

 

Battery: 9V battery, 006P or IEC6F22 or NEDA1604. 

3.8

 

Battery life: 150 hours for alkaline battery, and 80 hours for carbon battery. 

3.9

 

Dimension: 130mm (L) × 95mm (W) × 28mm (H). 

3.10

 

Weight: approx.240g (including battery).   

3.11

 

Accessories: manual, two pieces K-type TP01 thermocouple, 9V battery. 
 

4.

 

TECHNICAL DATA 

Accuracy adjustment environment: 23

±5

Thermocouple 

type 

Input range 

Resolution

Accuracy Remark 

-40

1760

 

1

 

-40

500

: ±(0.5%+2 )

 

500

℃~

1760

: ±(0.4%+2 )

 

T1-T2 

accurate: 

±

[0.3% 

(T1-T2)

2] 

Input 

impedance:  

1G

Adopts 
ITS-90 

thermometr

ic scale 

 

-20

1760

 

1

 

400

1800

 

1

 

400

℃~

800

: ±(0.5%+2 )

 

800

℃~

1000

: ±(0.4%+2 )

 

1000

℃~

1800

: ±(0.3%+2 )

 

-200

1000

 0.1

 

-200

℃~

0

: ±(0.3%+1 )

 

0

℃~

500

: ±(0.3%+2 )

 

-200

1370

 0.1

 

-200

0

: ±(0.3%+2 )

 

0

℃~

800

: ±(0.3%+1 )

 

800

℃~

1370

: ±(0.5%+2 )

 

-200

1200

 0.1

 

-200

0

: ±(0.3%+1 )

 

0

℃~

1200

: ±(0.3%+2 )

 

-200

400

 0.1

 

-200

0

: ±(0.3%+1 )

 

0

℃~

400

: ±(0.3%+2 )

 

-200

1300

 0.1

 

-200

0

: ±(0.5%+2 )

 

0

℃~

1300

: ±(0.3%+2 )

 

Note

 

1.

 

For obtaining an accurate measurement result, please warm up for 5minuetes before taking measurements. 

2.

 

The accuracy of above sheet doesn’t include the error of probe; please amend the accuracy according to the 
error of used probe. 

3.

 

The accuracy doesn’t include the inside temperature compensate sensor’s accuracy. 
Environmental temperature compensate range: -10

50

°

C, compensate value 

 0. 5

4.

 

Temperature coefficient: 

±

0.01%×range / 

°

0

°

C

18

°

C, 28

°

C

50

°

C). 

 

5.

 

PANEL DESCRIPTION 

 

LCD: main display shows temperature of T1, T2 and T1-T2. Auxiliary display shows the relative input 

voltage.  

 

[POWER]: turn ON/OFF the thermometer. 

 

[T1-T2]: choose temperature of T1, T2 or T1-T2. 

 

[

/

]: choose 

 or 

 temperature mode 

 

[MAX/MIN]: show the maximum and minimum 
temperature value. 

 

[HOLD]: press [HOLD] button to hold the current 
measurement value on the LCD. 

 

[LOG]: Memory function key   

 

T1: temperature input terminal, max input voltage 30V. 

 

T2: temperature input terminal, max input voltage 30V. 

 

OUTPUT: alarm signal output terminal.     

 

Symbol explanation 

Reviews: